Int Orthop. 2011 Aug;35(8):1131-7. doi: 10.1007/s00264-010-1077-6. Epub 2010 Jul 13.
Shell deformation of resurfacing and all-metal modular cups following press-fit implantation has been reported, but not for conventional metal-backed cups with polyethylene liners. The deformation of acetabular components with historical and thin polyethylene inserts after press-fit insertion was evaluated using a cadaveric model. All shells and liners deformed upon implantation. Following joint loading, shell pinch decreased from 0.32 to 0.22 mm (p = 0.019) and from 0.29 to 0.13 mm (p = 0.003) for the thin and thick liner groups, respectively. Liner pinch also decreased from 0.17 to 0.04 mm (p = 0.031) and from 0.06 to 0 mm (p = 0.103) for the thin and thick liner groups, respectively. There were no significant differences between the thin and thick liners. Liner deformation was influenced by the initial shell deformation and donor bone quality. Shell and liner pinch decreased following joint loading, suggesting a settling in effect.
翻修和全金属组合杯在压配植入后发生了外壳变形,这已有报道,但具有聚乙烯衬垫的传统金属背衬杯则没有发生过这种情况。使用尸体模型评估了压配植入后历史悠久且衬垫较薄的聚乙烯髋臼组件的髋臼部件变形情况。所有的外壳和衬垫在植入时都发生了变形。关节加载后,对于薄衬垫组,外壳压痕从 0.32 毫米减少到 0.22 毫米(p=0.019),从 0.29 毫米减少到 0.13 毫米(p=0.003);对于厚衬垫组,外壳压痕从 0.17 毫米减少到 0.04 毫米(p=0.031),从 0.06 毫米减少到 0 毫米(p=0.103)。薄衬垫组和厚衬垫组之间没有显著差异。衬垫变形受初始外壳变形和供体骨质量的影响。关节加载后,外壳和衬垫压痕减小,表明有稳定效果。