Rapid Commun Mass Spectrom. 2013 Jan 15;27(1):207-15. doi: 10.1002/rcm.6431.
The ratio of the measured abundance of (13)C-(18)O bonding CO(2) to its stochastic abundance, prescribed by the δ(13)C and δ(18)O values from a carbonate mineral, is sensitive to its growth temperature. Recently, clumped-isotope thermometry, which uses this ratio, has been adopted as a new tool to elucidate paleotemperatures quantitatively.
Clumped isotopes in CO(2) were measured with a small-sector isotope ratio mass spectrometer. CO(2) samples digested from several kinds of calcium carbonates by phosphoric acid at 25 °C were purified using both cryogenic and gas-chromatographic separations, and their isotopic composition (δ(13)C, δ(18)O, Δ(47), Δ(48) and Δ(49) values) were then determined using a dual-inlet Delta XP mass spectrometer.
The internal precisions of the single gas Δ(47) measurements were 0.005 and 0.02‰ (1 SE) for the optimum and the routine analytical conditions, respectively, which are comparable with those obtained using a MAT 253 mass spectrometer. The long-term variations in the Δ(47) values for the in-house working standard and the heated CO(2) gases since 2007 were close to the routine, single gas uncertainty while showing seasonal-like periodicities with a decreasing trend. Unlike the MAT 253, the Delta XP did not show any significant relationship between the Δ(47) and δ(47) values.
The Delta XP gave results that were approximately as precise as those of the MAT 253 for clumped-isotope analysis. The temporal stability of the Delta XP seemed to be lower, although an advantage of the Delta XP was that no dependency of δ(47) on Δ(47) was found.
由碳酸盐矿物的 δ(13)C 和 δ(18)O 值规定的测量的 (13)C-(18)O 键合 CO2 的丰度与随机丰度之比,对其生长温度敏感。最近,使用该比率的凝聚同位素测温法已被采用作为定量阐明古温度的新工具。
使用小扇区同位素质谱仪测量 CO2 中的凝聚同位素。用磷酸在 25°C 从几种碳酸钙中消化 CO2 样品,然后使用低温和气相色谱分离进行纯化,并使用双入口 Delta XP 质谱仪测定其同位素组成(δ(13)C、δ(18)O、Δ(47)、Δ(48)和 Δ(49)值)。
在最佳和常规分析条件下,单个气体 Δ(47)测量的内部精度分别为 0.005 和 0.02‰(1 SE),与使用 MAT 253 质谱仪获得的精度相当。自 2007 年以来,内部工作标准和加热 CO2 气体的 Δ(47)值的长期变化接近常规的单个气体不确定性,同时显示出具有下降趋势的季节性周期性。与 MAT 253 不同,Delta XP 没有显示 Δ(47)和 δ(47)值之间的任何显著关系。
Delta XP 给出的凝聚同位素分析结果与 MAT 253 的结果大致相同。尽管 Delta XP 的一个优势是没有发现 δ(47)对 Δ(47)的依赖性,但 Delta XP 的时间稳定性似乎较低。