Cadmium-zinc-telluride (CZT) solid-state detectors have been recently introduced into the field of myocardial perfusion imaging. The aim of this study was to prospectively compare the diagnostic performance of the CZT high-speed gamma camera (Discovery NM 530c) with that of the standard 3-head gamma camera in the same group of patients.
The study group consisted of 150 consecutive patients who underwent a 1-day stress-rest (99m)Tc-sestamibi or tetrofosmin imaging protocol. Image acquisition was performed first on a standard gamma camera with a 15-min scan time each for stress and for rest. All scans were immediately repeated on a CZT camera with a 5-min scan time for stress and a 3-min scan time for rest, using list mode. The correlations between the CZT camera and the standard camera for perfusion and function analyses were strong within narrow Bland-Altman limits of agreement. Using list mode analysis, image quality for stress was rated as good or excellent in 97% of the 3-min scans, and in 100% of the ≥4-min scans. For CZT scans at rest, similarly, image quality was rated as good or excellent in 94% of the 1-min scans, and in 100% of the ≥2-min scans.
The novel CZT camera provides excellent image quality, which is equivalent to standard myocardial single-photon emission computed tomography, despite a short scan time of less than half of the standard time.
碲锌镉(CZT)固态探测器最近已被引入心肌灌注成像领域。本研究的目的是前瞻性比较 CZT 高速伽马相机(Discovery NM 530c)与同组患者标准 3 头伽马相机的诊断性能。
研究组包括 150 例连续患者,他们在 1 天内接受了(99m)Tc- sestamibi 或四氮茂心肌灌注/功能成像。首先使用标准伽马相机以 15 分钟/帧的时间进行应激和静息扫描。所有扫描立即在 CZT 相机上重复,使用列表模式,应激扫描时间为 5 分钟,静息扫描时间为 3 分钟。在狭窄的 Bland-Altman 一致性界限内,CZT 相机与标准相机在灌注和功能分析方面的相关性很强。使用列表模式分析,在 3 分钟扫描中,97%的应激图像质量评为良好或优秀,在≥4 分钟扫描中,100%的图像质量评为良好或优秀。对于 CZT 静息扫描,同样,在 1 分钟扫描中,94%的图像质量评为良好或优秀,在≥2 分钟扫描中,100%的图像质量评为良好或优秀。
尽管扫描时间短至标准时间的一半以下,新型 CZT 相机仍能提供优质的图像质量,与标准单光子发射计算机断层扫描相当。