Cardiff School of Engineering, Cardiff University, Queen's Buildings, The Parade, Cardiff, CF24 3AA, UK.
Eur Spine J. 2013 Mar;22 Suppl 1(Suppl 1):S10-5. doi: 10.1007/s00586-012-2641-0. Epub 2013 Jan 4.
To compare the effectiveness of the Aspen, Aspen Vista, Philadelphia, Miami-J and Miami-J Advanced collars at restricting cervical spine movement in the sagittal, coronal and axial planes.
Nineteen healthy volunteers (12 female, 7 male) were recruited to the study. Collars were fitted by an approved physiotherapist. Eight ProReflex (Qualisys, Sweden) infrared cameras were used to track the movement of retro-reflective marker clusters placed in predetermined positions on the head and trunk. 3D kinematic data were collected during forward flexion, extension, lateral bending and axial rotation from uncollared to collared subjects. The physiological range of motion in the three planes was analysed using the Qualisys Track Manager System.
The Aspen and Philadelphia were significantly more effective at restricting flexion/extension than the Vista (p < 0.001), Miami-J (p < 0.001 and p < 0.01) and Miami-J Advanced (p < 0.01 and p < 0.05). The Aspen was significantly more effective at restricting rotation than the Vista (p < 0.001) and the Miami-J (p < 0.05). The Vista was significantly the least effective collar at restricting lateral bending (p < 0.001).
Our motion analysis study found the Aspen collar to be superior to the other collars when measuring restriction of movement of the cervical spine in all planes, particularly the sagittal and transverse planes, while the Aspen Vista was the least effective collar.
比较 Aspen、Aspen Vista、Philadelphia、Miami-J 和 Miami-J Advanced 颈托在限制颈椎矢状面、冠状面和轴向运动方面的效果。
招募了 19 名健康志愿者(12 名女性,7 名男性)参与研究。由经过认证的物理治疗师为参与者佩戴颈托。使用 8 个 ProReflex(Qualisys,瑞典)红外摄像机追踪放置在头部和躯干预定位置的反射标记集群的运动。在无颈托到佩戴颈托的情况下,从向前弯曲、伸展、侧屈和轴向旋转收集 3D 运动学数据。使用 Qualisys Track Manager 系统分析三个平面的生理运动范围。
Aspen 和 Philadelphia 比 Vista(p<0.001)、Miami-J(p<0.001 和 p<0.01)和 Miami-J Advanced(p<0.01 和 p<0.05)更有效地限制屈伸。Aspen 比 Vista(p<0.001)和 Miami-J(p<0.05)更有效地限制旋转。Vista 是限制侧屈效果最差的颈托(p<0.001)。
我们的运动分析研究发现,在测量颈椎在所有平面(尤其是矢状面和额状面)运动受限方面,Aspen 颈托优于其他颈托,而 Aspen Vista 是效果最差的颈托。