Anal Chem. 2013 Feb 19;85(4):2127-34. doi: 10.1021/ac3027542. Epub 2013 Jan 28.
Quantitative nontarget analysis is intended to provide a measurement of concentration of newly identified components in complex biological or environmental samples for which authentic or labeled standard do not exist. Electrospray ionization-high field asymmetric waveform ion mobility spectrometry-mass spectrometry (ESI-FAIMS-MS) has unique advantages that allowed us to develop a novel approach for quantification of nontarget analytes. In the nontarget analysis of urinary metabolites by ESI-FAIMS-MS, we find it practical and beneficial to analyze highly diluted urine samples. We show that urine extracts can be analyzed directly at very high dilutions (up to 20,000 times) by extending MS analysis times during slow FAIMS scanning. We explore the effects of sample dilution on ionization efficiency and ionization suppression in direct electrospray of complex sample matrixes. We consistently observe two distinct regimes in ESI operation related to the limited ionization capacity of this method. In the linear dynamic concentration range below the limiting ionization capacity, the analytical sensitivity of an analyte is constant and does not depend on matrix composition and concentration. Once the capacity of ESI is exceeded, all species exhibit log-log linearity in signal response. We show how quantification can be carried out using two different approaches, one for analytes which can be detected in the linear regime and another for those only detected in the suppression regime that overcomes the effects of ionization suppression. Our new insight into ionization suppression effects in ESI is of broad interest to anyone using ESI as an ionization technique for the MS analysis of complex samples.
定量非靶向分析旨在提供对复杂生物或环境样品中未识别成分浓度的测量,这些样品不存在真实或标记的标准品。电喷雾电离-高场非对称波形离子淌度谱-质谱(ESI-FAIMS-MS)具有独特的优势,使我们能够开发出一种用于非靶向分析物定量的新方法。在 ESI-FAIMS-MS 对尿代谢物的非靶向分析中,我们发现分析高度稀释的尿样是实用且有益的。我们通过在缓慢 FAIMS 扫描期间延长 MS 分析时间,展示了可以直接在非常高的稀释度(高达 20,000 倍)下分析尿提取物。我们探讨了样品稀释对复杂样品基质直接电喷雾中电离效率和电离抑制的影响。我们一致观察到与该方法的有限电离能力相关的 ESI 操作中的两个不同状态。在低于极限电离能力的线性动态浓度范围内,分析物的分析灵敏度是恒定的,与基质组成和浓度无关。一旦超过 ESI 的能力,所有物质的信号响应都呈对数线性关系。我们展示了如何使用两种不同的方法进行定量,一种用于可以在线性状态下检测到的分析物,另一种用于仅在抑制状态下检测到的分析物,该方法克服了电离抑制的影响。我们对 ESI 中电离抑制效应的新认识对任何使用 ESI 作为 MS 分析复杂样品的电离技术的人都具有广泛的兴趣。