三维弹性成像测量颈淋巴结体积:一项研究探讨可行性、准确性和可靠性。

Three-dimensional elastography for cervical lymph node volume measurements: a study to investigate feasibility, accuracy and reliability.

机构信息

Department of Health Technology and Informatics, Hong Kong Polytechnic University, Hung Hom, Kowloon, Hong Kong.

出版信息

Ultrasound Med Biol. 2013 Mar;39(3):396-406. doi: 10.1016/j.ultrasmedbio.2012.10.005. Epub 2013 Jan 11.

Abstract

This study investigated the feasibility of using three-dimensional (3-D) elastography in measuring cervical lymph node volume and compared the accuracy and reliability of 3-D elastography and 3-D grayscale ultrasound in measurement of ill-defined cervical nodes. Eighteen porcine lymph nodes from the neck were embedded in tissue-mimicking phantoms and scanned with the two ultrasound techniques. Ultrasound measurements were compared with the volume determined by water-displacement method to evaluate measurement accuracy. Inter-observer reproducibility and intra-observer repeatability of measurements were evaluated. Four patients with enlarged neck nodes were included to evaluate intra-observer repeatability of ultrasound measurements. Results demonstrated that lymph nodes that appeared ill-defined on grayscale ultrasound showed well-defined boundaries on elastography. 3-D elastography has higher measurement accuracy (84.2%), reproducibility (intraclass correlation coefficient, ICC = 0.909) and repeatability (ICC = 0.964-0.988) than does 3-D grayscale ultrasound (62.2%, 0.777 and 0.863-0.906 respectively). As a conclusion, 3-D elastography is accurate and reliable in volume measurement of ill-defined lymph nodes and has potential for accurate assessment of lymph node volume.

摘要

本研究探讨了三维(3-D)弹性成像在测量颈部淋巴结体积中的可行性,并比较了 3-D 弹性成像和 3-D 灰阶超声在测量边界不清的颈部淋巴结中的准确性和可靠性。将 18 个来自颈部的猪淋巴结嵌入组织模拟体模中,并使用两种超声技术进行扫描。将超声测量值与水置换法确定的体积进行比较,以评估测量的准确性。评估了测量的观察者间可重复性和观察者内重复性。纳入 4 例颈部淋巴结肿大的患者,以评估超声测量的观察者内重复性。结果表明,在灰阶超声上表现为边界不清的淋巴结在弹性成像上显示出明确的边界。3-D 弹性成像在测量准确性(84.2%)、可重复性(组内相关系数,ICC=0.909)和重复性(ICC=0.964-0.988)方面均优于 3-D 灰阶超声(分别为 62.2%、0.777 和 0.863-0.906)。总之,3-D 弹性成像在测量边界不清的淋巴结体积方面具有准确性和可靠性,具有准确评估淋巴结体积的潜力。

文献检索

告别复杂PubMed语法,用中文像聊天一样搜索,搜遍4000万医学文献。AI智能推荐,让科研检索更轻松。

立即免费搜索

文件翻译

保留排版,准确专业,支持PDF/Word/PPT等文件格式,支持 12+语言互译。

免费翻译文档

深度研究

AI帮你快速写综述,25分钟生成高质量综述,智能提取关键信息,辅助科研写作。

立即免费体验