Physics and Astronomy, University of British Columbia, Vancouver, BC V6T 1Z4, Canada. Medical Physics, BC Cancer Agency-Vancouver Centre, Vancouver, BC V5Z 4E6, Canada.
Phys Med Biol. 2013 Feb 21;58(4):749-70. doi: 10.1088/0031-9155/58/4/749. Epub 2013 Jan 16.
Four-dimensional volumetric modulated arc therapy (4D VMAT) is a treatment strategy for lung cancers that aims to exploit relative target and tissue motion to improve organ at risk (OAR) sparing. The algorithm incorporates the entire patient respiratory cycle using 4D CT data into the optimization process. Resulting treatment plans synchronize the delivery of each beam aperture to a specific phase of target motion. Stereotactic body radiation therapy treatment plans for 4D VMAT, gated VMAT, and 3D VMAT were generated on three patients with non-small cell lung cancer. Tumour motion ranged from 1.4-3.4 cm. The dose and fractionation scheme was 48 Gy in four fractions. A B-spline transformation model registered the 4D CT images. 4D dose volume histograms (4D DVH) were calculated from total dose accumulated at the maximum exhalation. For the majority of OARs, gated VMAT achieved the most radiation sparing but treatment times were 77-148% longer than 3D VMAT. 4D VMAT plan qualities were comparable to gated VMAT, but treatment times were only 11-25% longer than 3D VMAT. 4D VMAT's improvement of healthy tissue sparing can allow for further dose escalation. Future study could potentially adapt 4D VMAT to irregular patient breathing patterns.
四维容积调强弧形治疗(4D VMAT)是一种治疗肺癌的策略,旨在利用肿瘤和组织的相对运动来改善危及器官(OAR)的保护。该算法将整个患者呼吸周期的 4D CT 数据纳入到优化过程中。生成的治疗计划将每个射束孔径的传输与目标运动的特定相位同步。对三例非小细胞肺癌患者生成了 4D VMAT、门控 VMAT 和 3D VMAT 的立体定向体部放射治疗计划。肿瘤运动范围为 1.4-3.4cm。剂量和分割方案为 48Gy,共 4 次。B 样条变换模型对 4D CT 图像进行了配准。4D 剂量体积直方图(4D DVH)是从最大呼气时累积的总剂量计算得出的。对于大多数 OAR,门控 VMAT 实现了最大的放射保护,但治疗时间比 3D VMAT 长 77-148%。4D VMAT 计划质量与门控 VMAT 相当,但治疗时间仅比 3D VMAT 长 11-25%。4D VMAT 对健康组织保护的改善可以允许进一步提高剂量。未来的研究可能会使 4D VMAT 适应不规则的患者呼吸模式。