National Research Institute of Police Science, Kashiwa, Chiba 277-0882, Japan.
Anal Chem. 2013 Mar 5;85(5):2659-66. doi: 10.1021/ac303373u. Epub 2013 Feb 8.
A new method for sensitively and selectively detecting chemical warfare agents (CWAs) in air was developed using counter-flow introduction atmospheric pressure chemical ionization mass spectrometry (MS). Four volatile and highly toxic CWAs were examined, including the nerve gases sarin and tabun, and the blister agents mustard gas (HD) and Lewisite 1 (L1). Soft ionization was performed using corona discharge to form reactant ions, and the ions were sent in the direction opposite to the airflow by an electric field to eliminate the interfering neutral molecules such as ozone and nitrogen oxide. This resulted in efficient ionization of the target CWAs, especially in the negative ionization mode. Quadrupole MS (QMS) and ion trap tandem MS (ITMS) instruments were developed and investigated, which were movable on the building floor. For sarin, tabun, and HD, the protonated molecular ions and their fragment ions were observed in the positive ion mode. For L1, the chloride adduct ions of L1 hydrolysis products were observed in negative ion mode. The limit of detection (LOD) values in real-time or for a 1 s measurement monitoring the characteristic ions were between 1 and 8 μg/m(3) in QMS instrument. Collision-induced fragmentation patterns for the CWAs were observed in an ITMS instrument, and optimized combinations of the parent and daughter ion pairs were selected to achieve real-time detection with LOD values of around 1 μg/m(3). This is a first demonstration of sensitive and specific real-time detection of both positively and negatively ionizable CWAs by MS instruments used for field monitoring.
开发了一种新的方法,使用逆流导入常压化学电离质谱(MS)灵敏且选择性地检测空气中的化学战剂(CWA)。研究了四种挥发性和高毒性的 CWA,包括神经毒气沙林和塔崩,以及糜烂性毒剂芥子气(HD)和路易氏剂 1(L1)。采用电晕放电进行软电离以形成反应离子,并通过电场将离子沿与气流相反的方向输送,以消除臭氧和氮氧化物等干扰中性分子。这导致目标 CWA 的有效电离,特别是在负离子模式下。开发并研究了可在建筑物地板上移动的四极杆 MS(QMS)和离子阱串联 MS(ITMS)仪器。对于沙林、塔崩和 HD,在正离子模式下观察到质子化分子离子及其碎片离子。对于 L1,在负离子模式下观察到 L1 水解产物的氯加合物离子。在 QMS 仪器中,实时或用于 1 s 测量监测特征离子的检测限(LOD)值在 1 到 8 μg/m(3)之间。在 ITMS 仪器中观察到 CWA 的碰撞诱导碎裂模式,并选择母体和子离子对的最佳组合,以实现 LOD 值约为 1 μg/m(3)的实时检测。这是首次使用用于现场监测的 MS 仪器对正负可电离 CWA 进行灵敏和特异性实时检测的演示。