HUS Medical Imaging Center, Helsinki University Central Hospital, POB 340, Haartmaninkatu 4, 00290, Helsinki, Finland.
Pediatr Radiol. 2013 Jun;43(6):673-84. doi: 10.1007/s00247-012-2611-z. Epub 2013 Jan 23.
CT optimization has a special importance in children. Smaller body size accentuates the importance of patient positioning affecting both radiation dose and image quality.
To determine the effect of vertical positioning on organ dose, image noise and contrast in pediatric chest CT examination.
Chest scans of a pediatric 5-year anthropomorphic phantom were performed in different vertical positions (-6 cm to +5.4 cm) with a 64-slice CT scanner. Organ doses were measured with metal-oxide-semiconductor field-effect transistor (MOSFET) dosimeters. Image noise and contrast were determined from the CT number histograms corresponding to different tissues.
Significant changes in organ doses resulting from vertical positioning were observed, especially in radiosensitive anterior organs. The breast dose increased up to 16% and the thyroid dose up to 24% in lower positions. The noise was increased up to 45% relative to the centre position in the highest and lowest vertical positions, with a particular increase observed on the anterior and posterior sides, respectively. Off-centering also affected measured image contrast.
Vertical off-centering markedly affects organ doses and measured image-quality parameters in pediatric chest CT examination. Special attention should be given to correct patient centering when preparing patients for CT scans, especially when imaging children.
CT 优化在儿科中具有特殊的重要性。较小的体型强调了患者定位对辐射剂量和图像质量的影响。
确定在儿科胸部 CT 检查中垂直定位对器官剂量、图像噪声和对比的影响。
使用 64 层 CT 扫描仪对儿科 5 岁人体模型进行了不同垂直位置(-6cm 至+5.4cm)的胸部扫描。使用金属氧化物半导体场效应晶体管(MOSFET)剂量计测量器官剂量。从对应于不同组织的 CT 数直方图确定图像噪声和对比度。
观察到垂直定位导致器官剂量的显著变化,尤其是在前部敏感器官。在较低的位置,乳房剂量增加了 16%,甲状腺剂量增加了 24%。在最高和最低的垂直位置,与中心位置相比,噪声增加了 45%,在前部和后部分别观察到特别增加。偏离中心也会影响测量的图像对比度。
垂直偏离中心在儿科胸部 CT 检查中显著影响器官剂量和测量的图像质量参数。在为 CT 扫描准备患者时,特别是在为儿童成像时,应特别注意正确的患者定位。