全自动血液分析仪Sysmex XE - 5000用于腹水流式细胞术分析的评估。

Evaluation of the fully automated hematological analyzer Sysmex XE-5000 for flow cytometric analysis of peritoneal fluid.

作者信息

Lippi Giuseppe, Cattabiani Clarissa, Benegiamo Anna, Gennari Daniela, Pavesi Fernanda, Caleffi Alberta, Pipitone Silvia

机构信息

Laboratory of Clinical Chemistry and Hematology, Academic Hospital of Parma, Parma, Italy.

出版信息

J Lab Autom. 2013 Jun;18(3):240-4. doi: 10.1177/2211068212471833. Epub 2013 Jan 23.

Abstract

Although microscopy still represents the gold standard for cytometric analysis of peritoneal fluids, automated flow cytometry may improve throughput and accuracy. We evaluated the performance of total nucleated cell (TNC), white blood cell (WBC), polymorphonuclear cell (PMN), and mononuclear cell (MONO) counts of Sysmex XE-5000 on peritoneal fluids. The imprecision was excellent, being always lower than 11%, whereas linearity studies yielded correlation coefficients of 1.00 for all parameters. The carryover was always lower than 0.2%. The comparison between XE-5000 and microscopic analysis of 117 ascitic fluids yielded correlation coefficients always greater than 0.96, with mean biases <11/µL. The diagnostic accuracy versus manual microscopy was greater than that of XE-2100, especially at thresholds for septic ascites (100 versus 98% for ≥500 WBC/µL; 98 versus 93% for ≥250 PMN/µL). The correlation with manual microscopy for macrophages and mesothelial cell count was also higher for XE-5000 than for XE-2100 (0.63 versus 0.55). The results of this evaluation show optimal performance of XE-5000 for routine analysis of ascitic fluids, which are combined with the advantages of automated analysis such as high throughout, shortened turnaround time, no need of sample preparation and trained staff, reduced sample volume, and less likelihood of transcriptional errors.

摘要

尽管显微镜检查仍是腹水细胞计数分析的金标准,但自动流式细胞术可提高检测通量和准确性。我们评估了Sysmex XE-5000对腹水进行总核细胞(TNC)、白细胞(WBC)、多形核细胞(PMN)和单核细胞(MONO)计数的性能。其不精密度极佳,始终低于11%,而线性研究得出所有参数的相关系数均为1.00。携带污染率始终低于0.2%。对117份腹水进行XE-5000检测与显微镜分析的比较,得出相关系数始终大于0.96,平均偏差<11/µL。与手动显微镜检查相比,XE-5000的诊断准确性高于XE-2100,尤其是在诊断感染性腹水的阈值方面(白细胞≥500/µL时分别为100%对98%;多形核细胞≥250/µL时分别为98%对93%)。与手动显微镜检查相比,XE-5000对巨噬细胞和间皮细胞计数的相关性也高于XE-2100(分别为0.63对0.55)。本评估结果显示,XE-5000在腹水常规分析中表现优异,兼具自动分析的诸多优势,如高通量、缩短周转时间、无需样本制备及专业技术人员、减少样本量以及降低转录错误的可能性。

文献AI研究员

20分钟写一篇综述,助力文献阅读效率提升50倍。

立即体验

用中文搜PubMed

大模型驱动的PubMed中文搜索引擎

马上搜索

文档翻译

学术文献翻译模型,支持多种主流文档格式。

立即体验