[低千伏设置在下肢计算机断层扫描静脉造影中的应用价值]

[Usefulness of low kilovoltage settings in computed tomography venography of lower limbs].

作者信息

Kuramochi Kenji, Ogawa Yasuyoshi, Chikaraishi Kousuke, Tateishi Kiyoko, Yoshikawa Tatsuo

机构信息

Imaging Center, St. Marianna University School of Medicine Hospital.

出版信息

Nihon Hoshasen Gijutsu Gakkai Zasshi. 2013 Jan;69(1):85-91. doi: 10.6009/jjrt.2013_jsrt_69.1.85.

Abstract

It has been reported that a reduction in tube kilovoltage during computed tomography (CT) angiography results in an average reduction of the effective radiation dose. Furthermore, a lower kilovoltage has been shown as a technique dose. However, there is no fundamental data in a low-kilovoltage protocol for CT venography. Thus, the purpose of this study was to investigate contrast enhancement, image noise, and radiation exposure with lower kilovoltage on CT images scanned using phantom of lower limbs and clinical CT images. In order to grasp the effective energy in each tube voltage of the equipment used, we determined the half-value layer using aluminum attenuation coefficient. The phantom of the lower was sealed with contrast agent that was adjusted in various CT values. We scanned this phantom at 80 kVp, 100 kVp, and 120 kVp settings, and evaluated the changes in CT value. We also compared CT values, CTDIvol, contrast enhancement, and radiation exposure with 100 kVp and 120 kVp in patients with suspected pulmonary embolism or deep venous thrombosis. We found the CT value increased 30 HU with 100 kVp settings, and contrast was also improved. A reduction of radiation exposure without deterioration of image quality would be possible by lowering the kilovoltage setting in CT venography.

摘要

据报道,在计算机断层扫描(CT)血管造影期间降低管电压会导致有效辐射剂量平均降低。此外,较低的管电压已被证明是一种技术剂量。然而,对于CT静脉造影的低管电压方案,尚无基础数据。因此,本研究的目的是使用下肢模型和临床CT图像,研究在较低管电压下CT图像的对比增强、图像噪声和辐射暴露情况。为了掌握所用设备各管电压下的有效能量,我们使用铝衰减系数确定了半值层。下肢模型用在各种CT值下调节好的造影剂密封。我们在80 kVp、100 kVp和120 kVp设置下扫描该模型,并评估CT值的变化。我们还比较了疑似肺栓塞或深静脉血栓形成患者在100 kVp和120 kVp时的CT值、容积CT剂量指数(CTDIvol)、对比增强和辐射暴露情况。我们发现,在100 kVp设置下CT值增加了30 HU,对比也得到了改善。在CT静脉造影中降低管电压设置,有可能在不降低图像质量的情况下减少辐射暴露。

文献检索

告别复杂PubMed语法,用中文像聊天一样搜索,搜遍4000万医学文献。AI智能推荐,让科研检索更轻松。

立即免费搜索

文件翻译

保留排版,准确专业,支持PDF/Word/PPT等文件格式,支持 12+语言互译。

免费翻译文档

深度研究

AI帮你快速写综述,25分钟生成高质量综述,智能提取关键信息,辅助科研写作。

立即免费体验