The COBAS Elecsys PTH(1-84) assay is a novel, electro-chemiluminescence immunoassay that exclusively measures full-length parathyroid hormone (PTH). The aim of this study is to compare the automated biointact Elecsys PTH(1-84) assay with four contemporary, iPTH assays in chronic kidney disease (CKD) patients.
We compared the Elecsys PTH(1-84) assay with four iPTH assays (Siemens ADVIA Centaur, Ortho Clinical Diagnostics (OCD) VITROS, Beckman Access2, Abbott ARCHITECT) in the measurement of PTH in 83 local CKD patients. Majority of the patients (44) had CKD but were not on dialysis, 15 were on hemodialysis, 15 were on peritoneal dialysis, and 9 were post-renal transplant. The precision performance and correlation of the assays were determined. PTH(1-84) concentrations were correlated with calcium, phosphate, alkaline phosphatase, hemoglobin, HbA1c and lipid concentrations.
The Elecsys PTH(1-84) assay showed comparable precision and good correlation with the iPTH assays. Although the four different iPTH assays correlated well with each other, there was significant discrepancy among assays. The discrepancy among assays increased with increasing PTH concentrations. The ADVIA Centaur and ARCHITECT assays measured significantly higher PTH concentrations than the VITROS and Access2 assays. PTH(1-84) showed a positive association with phosphate and alkaline phosphatase and an inverse association with HbA1c. There was no significant association with lipid concentrations.
The third generation Elecsys PTH(1-84) assay had comparable precision performance and correlated well with second generation iPTH assays. However, significant discrepancy was found among the four iPTH assays in measuring iPTH in CKD patients.
COBAS Elecsys PTH(1-84)检测法是一种新型的电化学发光免疫分析法,专门用于测量全长甲状旁腺激素(PTH)。本研究旨在比较全自动生物活性 Elecsys PTH(1-84)检测法与四种当代 iPTH 检测法在慢性肾脏病(CKD)患者中的应用。
我们比较了 Elecsys PTH(1-84)检测法与四种 iPTH 检测法(Siemens ADVIA Centaur、Ortho Clinical Diagnostics (OCD) VITROS、Beckman Access2、Abbott ARCHITECT)在 83 名本地 CKD 患者的 PTH 测量中的应用。大多数患者(44 名)患有 CKD 但未接受透析,15 名患者接受血液透析,15 名患者接受腹膜透析,9 名患者为肾移植后患者。测定了各检测法的精密度性能和相关性。将 PTH(1-84)浓度与钙、磷、碱性磷酸酶、血红蛋白、HbA1c 和血脂浓度进行相关分析。
Elecsys PTH(1-84)检测法显示出与 iPTH 检测法相当的精密度和良好的相关性。虽然四种不同的 iPTH 检测法彼此之间相关性良好,但各检测法之间存在显著差异。随着 PTH 浓度的增加,各检测法之间的差异也随之增加。ADVIA Centaur 和 ARCHITECT 检测法测量的 PTH 浓度明显高于 VITROS 和 Access2 检测法。PTH(1-84)与磷和碱性磷酸酶呈正相关,与 HbA1c 呈负相关。与血脂浓度无显著相关性。
第三代 Elecsys PTH(1-84)检测法具有相当的精密度性能,与第二代 iPTH 检测法相关性良好。然而,在测量 CKD 患者的 iPTH 时,四种 iPTH 检测法之间存在显著差异。