Vidgren M, Paronen P, Vidgren P, Vainio P, Nuutinen J
Department of Pharmaceutical Technology, University of Kuopio, Finland.
Acta Pharm Nord. 1990;2(1):3-10.
Deposition of 99mTc-labelled particles of disodium cromoglycate in the respiratory tract was evaluated after inhalation from two dry powder inhalers. The inhalers tested were a new multiple dose powder device (Chiesi Farmaceutici) and the Rotahaler powder device (Glaxo). Fractional deposition in the whole lung area, the upper respiratory passages and the stomach, as well as in the inhaler, was measured using a gamma camera. Ten healthy volunteers participated in the inhalation test. The mean, whole lung deposition of the inhaled disodium cromoglycate particles was about 9 per cent of the dose for the both devices. The deposition patterns of the inhaled drug doses in the lung area given by the devices were, however, clearly different. The Chiesi powder inhaler was more efficient in dispersing the drug particles into the therapeutically important alveolar regions of the lungs. The fraction of the dose retained in the inhaler was significantly higher for the Rotahaler than for the Chiesi powder device. On the other hand, the deposition in the upper passages was somewhat larger for the Chiesi device than for the Rotahaler. According to the deposition results of this study, a new multiple dose powder inhaler can be useful in inhalation therapy.
从两种干粉吸入器吸入后,评估了99mTc标记的色甘酸钠颗粒在呼吸道中的沉积情况。所测试的吸入器一种是新型多剂量粉末装置(基耶西制药公司),另一种是旋转吸入器粉末装置(葛兰素公司)。使用γ相机测量了在整个肺部区域、上呼吸道、胃部以及吸入器中的分数沉积。十名健康志愿者参与了吸入测试。两种装置吸入的色甘酸钠颗粒的平均全肺沉积量约为剂量的9%。然而,这两种装置在肺部区域给予的吸入药物剂量的沉积模式明显不同。基耶西粉末吸入器在将药物颗粒分散到肺部具有治疗意义的肺泡区域方面更有效。旋转吸入器中保留的剂量分数明显高于基耶西粉末装置。另一方面,基耶西装置在上呼吸道中的沉积比旋转吸入器稍大。根据本研究的沉积结果,一种新型多剂量粉末吸入器可用于吸入治疗。