Department of Bioengineering, University of Pennsylvania, Philadelphia, PA 19104, USA.
J Allergy Clin Immunol. 2004 Oct;114(4):863-8. doi: 10.1016/j.jaci.2004.07.017.
Adherence to inhaled steroid regimens is frequently poor. Finding ways to improve adherence depends on the ability to measure time and date of inhaler use reliably and to detect deliberate dose dumping. There is no such monitor for the popular new dry powder inhalers.
To develop and test an electronic monitor for a dry powder inhaler that will provide information on the time and date of use.
An electronic adherence monitor for the Advair Diskus dry powder inhaler was developed and tested. In this device, inhaler use is determined by detecting and recording the motion of the drug delivery lever in the inhaler with a magnetic sensor. An additional electronic interface and software were also developed to allow the adherence data to be uploaded to a computer for display and analysis.
System and reliability tests involving multiple-day and repeated-use tests of the adherence monitor demonstrate the overall performance and reliability of the device, and specifically its ability to record the time and date of dose delivery. In the repeated-use test, 300 successive actuations of the dose delivery lever were correctly sensed and recorded without error.
The simple-to-use, low-cost, reusable adherence monitor accurately records time and date of inhaler use and thus allows clinical monitoring and adherence studies in patients using the Advair Diskus dry powder inhaler. The same technology should be adaptable to other dry powder inhalers, including the Pulmicort Tubuhaler and the Symbicort Tubuhaler.
吸入性类固醇治疗方案的依从性常常较差。找到提高依从性的方法取决于能否可靠地测量吸入器使用的时间和日期,并检测故意倾倒剂量的情况。对于流行的新型干粉吸入器,尚无此类监测器。
开发并测试一种用于干粉吸入器的电子监测器,该监测器将提供使用时间和日期的信息。
开发并测试了一种用于Advair Diskus干粉吸入器的电子依从性监测器。在该装置中,通过使用磁传感器检测并记录吸入器中药物输送杆的运动来确定吸入器的使用情况。还开发了一个额外的电子接口和软件,以便将依从性数据上传到计算机进行显示和分析。
涉及依从性监测器多日和重复使用测试的系统和可靠性测试证明了该装置的整体性能和可靠性,特别是其记录剂量输送时间和日期的能力。在重复使用测试中,剂量输送杆连续300次启动均被正确检测并记录,无错误发生。
这种使用简单、成本低廉、可重复使用的依从性监测器能准确记录吸入器的使用时间和日期,从而可用于对使用Advair Diskus干粉吸入器的患者进行临床监测和依从性研究。相同技术应可适用于其他干粉吸入器,包括普米克都保和信必可都保。