Department of Neurosurgery, Erasmus University Medical Centre, 's Gravendijkwal 230, Rotterdam 3000 CA, The Netherlands.
J Neurol Neurosurg Psychiatry. 2013 Jul;84(7):813-7. doi: 10.1136/jnnp-2012-302935. Epub 2013 Feb 13.
In treating idiopathic normal pressure hydrocephalus (INPH) with a shunt there is always a risk of underdrainage or overdrainage. The hypothesis is tested whether patients treated using an adjustable valve preset at the highest opening pressure leads to comparable good clinical results with less subdural effusions than in a control group with an opening pressure preset at a low pressure level.
A multicentre prospective randomised trial was performed on a total of 58 patients suspected of INPH. Thirty patients were assigned to (control) group 1 and received a Strata shunt (Medtronic, Goleta, USA) with the valve preset at a performance level (PL) of 1.0, while 28 patients were assigned to group 2 and received a Strata shunt with the valve preset at PL 2.5. In this group the PL was allowed to be lowered until improvement or radiological signs of overdrainage were met.
Significantly more subdural effusions were observed in the improved patients of group 1. There was no statistically significant difference in improvement between both groups overall.
On the basis of this multicentre prospective randomised trial it is to be recommended to treat patients with INPH with a shunt with an adjustable valve, preset at the highest opening pressure and lowered until clinical improvement or radiological signs of overdrainage occur although slower improvement and more shunt adjustments might be the consequence.
在使用分流器治疗特发性正常压力脑积水(INPH)时,始终存在引流不足或过度引流的风险。本研究旨在检验使用预设在最高开放压力的可调式阀门治疗的患者,是否与预设在较低压力水平的开放式阀门的对照组相比,具有更好的临床效果且更少发生硬膜下积液。
共对 58 例疑似 INPH 的患者进行了多中心前瞻性随机试验。30 例患者被分配到(对照组)1 组,接受 Strata 分流器(美敦力,加利福尼亚州,美国)治疗,阀门预设为性能水平(PL)1.0,而 28 例患者被分配到 2 组,接受 Strata 分流器治疗,阀门预设为 PL 2.5。在该组中,允许将 PL 降低到改善或出现影像学提示过度引流的程度。
1 组中改善的患者中观察到更多的硬膜下积液。两组之间的整体改善程度无统计学差异。
基于这项多中心前瞻性随机试验,建议使用预设在最高开放压力的可调式阀门治疗 INPH 患者,并在出现临床改善或影像学提示过度引流时降低阀门压力,尽管这可能导致改善更慢和更多的分流器调整。