J Glaucoma. 2014 Dec;23(9):633-7. doi: 10.1097/IJG.0b013e318285fefd.
To assess the accuracy of a rebound tonometer (RT) in patients with glaucoma, comparing the measurements with those obtained by a noncontact tonometer (NCT) and a Goldmann applanation tonometer (GAT).
A total of 71 patients (142 eyes) with glaucoma from our hospital-based glaucoma clinic were included in the study. Intraocular pressure (IOP) was measured with RT, NCT, and GAT with an interval of 10 minutes, in between the readings obtained by the different tonometers.
In eyes with normal IOP, the IOP measured by RT is 1 to 3 mm Hg higher than that by NCT and 2 to 3 mm Hg higher than that by GAT. There was a trend toward greater difference when the IOP values were higher (>21 mm Hg). Mean IOP readings recorded by RT were higher than IOP recorded by NCT and GAT. A statistically significant tendency to an overestimation of IOP with RT as compared with GAT was found.
IOP readings obtained by an I-care tonometer in our study have shown reasonable concordance with GAT. Overestimation of IOP measured by RT as compared with GAT is directly proportional to the IOP. As the I-care tonometer measured IOP with good accuracy, it may be considered as an appropriate method for clinical use in normal subjects and glaucoma patients.
评估回弹眼压计(RT)在青光眼患者中的准确性,将其测量结果与非接触眼压计(NCT)和压平眼压计(GAT)的测量结果进行比较。
本研究共纳入我院青光眼门诊的 71 例(142 只眼)青光眼患者。在不同眼压计的读数之间,每隔 10 分钟用 RT、NCT 和 GAT 测量一次眼压(IOP)。
在正常眼压眼,RT 测量的 IOP 比 NCT 高 1 至 3mmHg,比 GAT 高 2 至 3mmHg。当眼压值较高(>21mmHg)时,差异有增大的趋势。RT 记录的平均眼压读数高于 NCT 和 GAT 记录的眼压读数。与 GAT 相比,RT 存在显著的眼压高估趋势。
在本研究中,I-care 眼压计的眼压读数与 GAT 具有良好的一致性。与 GAT 相比,RT 测量的眼压高估程度与眼压值成正比。由于 I-care 眼压计具有良好的准确性,因此可考虑将其作为正常受试者和青光眼患者临床应用的合适方法。