To investigate the feasibility and to compare three devices measuring intraocular pressure (IOP) in mucopolysaccharidosis patients (MPS): iCare rebound tonometer (RT), Perkins applanation tonometer (PAT) and ocular response analyzer (ORA).
MPS patients who underwent at least two examinations out of: RT, PAT and ORA at the same visit were identified and retrospectively analyzed in this study.
17 patients fulfilled the inclusion criterion. In all 17 patients IOP measurements were performed with RT (34 eyes) and ORA (33 eyes), while PAT measurement was possible in only 12 (24 eyes) patients. The RT, corneal-compensated intraocular pressure (IOPcc) and Goldmann-correlated intraocular pressure (IOPg) differed relevantly from IOP assessed with PAT. Corneal clouding in MPS patients correlated positively with PAT, RT and IOPg (r = 0.3, 0.5, and 0.5 respectively), but not with IOPcc (r = 0.07). The MPS-related corneal clouding correlated positively with biomechanical corneal parameters assessed with ORA: corneal hysteresis (r = 0.77) and corneal resistance factor (r = 0.77) either.
RT and ORA measurements were tolerated better than applanation tonometry in MPS patients. IOP measurements assessed with RT and ORA differed relevantly from PAT. Corneal-compensated IOP assessed with ORA seems to be less affected by the MPS-related corneal clouding than applanation or rebound tonometry. RT and ORA measurements should be preferred for IOP assessment in patients with MPS.
研究在黏多糖贮积症(MPS)患者中使用三种测量眼压(IOP)的设备的可行性并进行比较:iCare回弹眼压计(RT)、帕金斯压平眼压计(PAT)和眼反应分析仪(ORA)。
本研究纳入并回顾性分析了在同一次就诊中接受了RT、PAT和ORA中至少两项检查的MPS患者。
17例患者符合纳入标准。所有17例患者均使用RT(34只眼)和ORA(33只眼)进行了眼压测量,而仅12例(24只眼)患者可行PAT测量。RT测量的角膜补偿眼压(IOPcc)和Goldmann相关眼压(IOPg)与PAT评估的眼压有显著差异。MPS患者的角膜混浊与PAT、RT和IOPg呈正相关(r分别为0.3、0.5和0.5),但与IOPcc无关(r = 0.07)。MPS相关的角膜混浊也与ORA评估的角膜生物力学参数呈正相关:角膜滞后(r = 0.77)和角膜阻力因子(r = 0.77)。
在MPS患者中,RT和ORA测量比压平眼压测量耐受性更好。RT和ORA评估的眼压测量值与PAT有显著差异。与压平眼压测量或回弹眼压测量相比,ORA评估的角膜补偿眼压似乎受MPS相关角膜混浊的影响较小。在MPS患者中,应首选RT和ORA测量来评估眼压。