Suppr超能文献

亚可见颗粒如何变得不可见——折射率在蛋白质颗粒分析中的相关性。

How subvisible particles become invisible-relevance of the refractive index for protein particle analysis.

机构信息

Coriolis Pharma, 82152 Martinsried-Munich, Germany.

出版信息

J Pharm Sci. 2013 May;102(5):1434-46. doi: 10.1002/jps.23479. Epub 2013 Mar 5.

Abstract

The aim of the present study was to quantitatively assess the relevance of transparency and refractive index (RI) on protein particle analysis by the light-based techniques light obscuration (LO) and Micro-Flow Imaging (MFI). A novel method for determining the RI of protein particles was developed and provided an RI of 1.41 for protein particles from two different proteins. An increased RI of the formulation by high protein concentration and/or sugars at pharmaceutically relevant levels was shown to lead to a significant underestimation of the subvisible particle concentration determined by LO and MFI. An RI match even caused particles to become "invisible" for the system, that is, not detectable anymore by LO and MFI. To determine the influence of formulation RI on particle measurements, we suggest the use of polytetrafluoroethylene (PTFE) particles to test a specific formulation for RI effects. In case of RI influences, we recommend also using a light-independent technique such as resonant mass measurement (RMM) (Archimedes) for subvisible particle analysis in protein formulations.

摘要

本研究的目的是定量评估光散射(LO)和微流成像(MFI)等基于光的技术中透明度和折射率(RI)对蛋白质颗粒分析的相关性。开发了一种用于确定蛋白质颗粒 RI 的新方法,该方法为两种不同蛋白质的蛋白质颗粒提供了 1.41 的 RI。结果表明,制剂的 RI 随着蛋白质浓度和/或在药学相关水平的糖的增加而增加,这会导致 LO 和 MFI 测定的亚可见颗粒浓度的严重低估。RI 匹配甚至会导致颗粒对系统“不可见”,即 LO 和 MFI 无法再检测到。为了确定制剂 RI 对颗粒测量的影响,我们建议使用聚四氟乙烯(PTFE)颗粒来测试特定制剂的 RI 影响。在存在 RI 影响的情况下,我们还建议使用光独立技术(如谐振质量测量(RMM)(阿基米德))来分析蛋白质制剂中的亚可见颗粒。

文献AI研究员

20分钟写一篇综述,助力文献阅读效率提升50倍。

立即体验

用中文搜PubMed

大模型驱动的PubMed中文搜索引擎

马上搜索

文档翻译

学术文献翻译模型,支持多种主流文档格式。

立即体验