J Chromatogr A. 2013 Apr 12;1285:148-58. doi: 10.1016/j.chroma.2013.01.080. Epub 2013 Feb 17.
Previous reports have highlighted the influence of the properties of the mobile phase flow rate on the column performance achieved in supercritical fluid chromatography (SFC). In SFC both the mass and the volumetric flow rates have unique influences on the chromatographic performance and the determination of their exact values is critical. It is well understood that the mass flow rate stays constant along an SFC system whereas the volumetric flow rate may vary considerably, but the extent of these variations and the role of the individual operating parameters in influencing these variations have not been clearly reported yet. The factors that control the mass and the volumetric flow rates in an SFC system are discussed and the possible extent of variations of these flow rates under different operating pressures and temperatures are demonstrate quantitatively.
先前的报告强调了流动相流速特性对超临界流体色谱(SFC)中实现的柱性能的影响。在 SFC 中,质量和体积流速都对色谱性能有独特的影响,因此确定它们的确切值至关重要。人们清楚地知道,在 SFC 系统中,质量流速保持恒定,而体积流速可能会有很大变化,但这些变化的程度以及各个操作参数在影响这些变化中的作用尚未得到明确报道。本文讨论了控制 SFC 系统中质量和体积流速的因素,并定量地展示了在不同操作压力和温度下这些流速可能发生的变化程度。