University of Tennessee, Knoxville, TN 37996, USA.
J Chromatogr A. 2013 Dec 13;1320:130-7. doi: 10.1016/j.chroma.2013.10.041. Epub 2013 Oct 22.
This work demonstrates the possible advantages and the challenges of accurate on-line measurements of the CO2 mass flow rate during supercritical fluid chromatography (SFC) operations. Only the mass flow rate is constant along the column in SFC. The volume flow rate is not. The critical importance of accurate measurements of mass flow rates for the achievement of reproducible data and the serious difficulties encountered in supercritical fluid chromatography for its assessment were discussed earlier based on the physical properties of carbon dioxide. In this report, we experimentally demonstrate the problems encountered when performing mass flow rate measurements and the gain that can possibly be achieved by acquiring reproducible data using a Coriolis flow meter. The results obtained show how the use of a highly accurate mass flow meter permits, besides the determination of accurate values of the mass flow rate, a systematic, constant diagnosis of the correct operation of the instrument and the monitoring of the condition of the carbon dioxide pump.
本文展示了在超临界流体色谱(SFC)操作过程中对 CO2 质量流率进行准确在线测量的可能优势和挑战。在 SFC 中,只有质量流率沿柱是恒定的。体积流率则不是。基于二氧化碳的物理性质,我们之前曾讨论过准确测量质量流率对于实现可重现数据的至关重要性,以及在超临界流体色谱中评估其的严重困难。在本报告中,我们通过实验演示了在进行质量流率测量时遇到的问题,以及使用科里奥利流量计获得可重现数据可能带来的收益。所得结果表明,除了确定准确的质量流率值外,使用高精度质量流量计还可以系统地、持续地诊断仪器的正确运行情况,并监测二氧化碳泵的状况。