Blomquist P H, Gross R L, Koch D D
Cullen Eye Institute, Department of Ophthalmology, Baylor College of Medicine, Houston, Tex. 77030.
Ophthalmic Surg. 1990 Mar;21(3):223-6.
A neodymium: YAG laser operating in the thermal mode was used to irradiate isolated intraocular lenses (IOLs) and to perform transscleral cyclophotocoagulation on pseudophakic autopsy eyes to investigate the potential damage to IOL haptics such irradiation may cause. In the isolated IOLs, 70 mJ of energy deformed and partially melted both polymethylmethacrylate (PMMA) and polypropylene haptics. One of the capsular-fixated PC-IOL haptics in an autopsy eye partially melted when irradiated with the maximum energy level (8.8 J), with the aiming beam focused 1 mm posterior to the limbus and maximal posterior focus offset.