Effect of anemia on pulse oximeter accuracy at low saturation.

作者信息

Severinghaus J W, Koh S O

机构信息

Department of Anesthesia, University of California, San Francisco 94143-0542.

出版信息

J Clin Monit. 1990 Apr;6(2):85-8. doi: 10.1007/BF02828282.

Abstract

A retrospective evaluation of simultaneous tests of oximeters of various manufacturers in volunteer subjects disclosed greater errors at low saturations in subjects with low hemoglobin (Hb) concentrations. Forty-three pulse oximeters of 12 manufacturers studied over a period of 10 months showed that, at a mean arterial oxygen saturation (SaO2) level of 54.5%, as Hb concentration fell, average pulse oximeter (SpO2) bias increased approximately linearly from 0 at Hb greater than 14 g/dl to about -14% at 8 less than Hb less than 9 g/dl. At SaO2 = 53.6%, the mean bias (SaO2--SpO2) of 13 oximeters of 5 manufacturers averaged -15.0% (n = 43) in a subject with Hb = 8 g/dl, but -6.4% (n = 390) in nonanemic subjects. The additional bias in the anemic subject increased with desaturation. It was 0.13% at SaO2 = 98.5% (n = 13), -1.31% at 87.5% (n = 38), -2.71% at 75.1% (n = 38), -5.18% at 61.3% (n = 26), and -9.95% at 53.6% (n = 41); n is the product of the number of oximeters and number of tests in each saturation range. The instruments that showed the greatest errors at low saturations in nonanemic subjects also showed the greatest additional errors associated with anemia (the range between manufacturers of anemic incremental error at about 53% being from -3.2 to -14.5%) and conformed well to the relationship bias (anemic) = 1.35 x bias (normal) -8.18% (r = 0.94; Sy.x = 3.3%). The error due to anemia was zero at 97% SaO2 and became evident when SaO2 fell below 75%.

摘要

文献检索

告别复杂PubMed语法,用中文像聊天一样搜索,搜遍4000万医学文献。AI智能推荐,让科研检索更轻松。

立即免费搜索

文件翻译

保留排版,准确专业,支持PDF/Word/PPT等文件格式,支持 12+语言互译。

免费翻译文档

深度研究

AI帮你快速写综述,25分钟生成高质量综述,智能提取关键信息,辅助科研写作。

立即免费体验