Accuracy of dual-energy radiographic absorptiometry of the lumbar spine: cadaver study.

作者信息

Ho C P, Kim R W, Schaffler M B, Sartoris D J

机构信息

Department of Radiology, University of California, San Diego.

出版信息

Radiology. 1990 Jul;176(1):171-3. doi: 10.1148/radiology.176.1.2353087.

Abstract

Dual-energy radiographic absorptiometry (DRA) was used to measure the bone mineral content and area density of lumbar vertebrae (L2-L3) in 11 cadavers. These data were subsequently compared with measured ash content and density. Excellent correlation was obtained between bone mineral content measured with DRA and ash weight (r = .963, P less than .0001). The accuracy error in determining mineral content in lumbar vertebrae with DRA was about 9%. In addition, strong correlation was observed between bone mineral density measured with DRA and ash density (r = .881, P less than .0001).