Guérin J L, Favre G, Raynal G, Soula G
Département de Biologie Clinique, Centre Claudius-Régaud, Toulouse.
Ann Biol Clin (Paris). 1990;48(3):167-70.
The authors have adapted a turbidimetric assay on centrifugal Cobas Fara analyser for cerebrospinal fluid proteins determinations. The later method is proposed by SFBC Protein Analysis Committee using benzethonium chloride as monoreagent. This adaptation is more sensitive than manual method. The linearity range is greater (0.08-1.85 g/l instead of 0.2-1.2 g/l). Sample assay and time of analysis were reduced by 90 p. cent.