Zaránd P, Polgár I, Katona E
Oncoradiological Centre, Emil Weil Hospital, Budapest, Hungary.
Radiobiol Radiother (Berl). 1990;31(2):151-4.
The sensitivity vs. dose characteristics of two TL readers (Harshaw 2000 A + B and NHZ-203C) was compared both in the range of small doses (10 microGy to 100 mGy) as well as in the radiation therapy level by using individually calibrated LiF:Mg, Ti detectors manufactured in Poland. The performance of both systems can be predicted by a three parameter theoretical model. The very good reproducibility of the NHZ-203C reader (sigma/D = 0.6%) is important in radiation therapy while the high sensitivity (the lowest detectable dose limit = 3 S.D. of the background = 27 microGy) is useful for radiation protection measurements.
使用波兰制造的经过单独校准的LiF:Mg, Ti探测器,在小剂量范围(10微戈瑞至100毫戈瑞)以及放射治疗剂量水平下,比较了两台热释光(TL)读数器(Harshaw 2000 A + B和NHZ - 203C)的灵敏度与剂量特性。两个系统的性能可以通过一个三参数理论模型来预测。NHZ - 203C读数器具有非常好的重现性(标准差/剂量=0.6%),这在放射治疗中很重要,而其高灵敏度(最低可探测剂量限值=背景值的3倍标准差=27微戈瑞)则对辐射防护测量很有用。