Kojima Shinya, Hirata Masami, Shinohara Hiroyuki, Ueno Eiko
Department of Radiology, Tokyo Women's Medical University Medical Center East, Arakawa-ku, Tokyo 116-8567, Japan.
Radiol Phys Technol. 2013 Jul;6(2):375-84. doi: 10.1007/s12194-013-0211-8. Epub 2013 Apr 11.
In follow-up brain magnetic resonance imaging (MRI), precise reproducibility of the scan prescription is important so that over- or underestimating changes in volumes of clinical interest is prevented. (The scan prescription is defined as the location and orientation of the head with respect to the scan planes of the three-dimensional MRI matrix.) In this study, the misregistration between the original and a second scan was calculated in the case of both manual positioning and automatic positioning. These calculations were carried out both for a healthy volunteer scanned repeatedly and, in a retrospective study, for 225 patients who had an original and at least one follow-up scan. The effects of the scan operator being the same for both scans or being different were also examined. A commercially available 1.5 Tesla MRI system and a six-element head-array coil were employed in all of the imaging. The reproducibility of the scan prescription was determined by the registration of the original scan image to the follow-up scan image by use of the Fourier phase correlation method. Our results showed that (1) the reproducibility by automatic positioning was superior to that by manual positioning (p < 0.05), and (2) there was no significant difference in the results between when the operator was the same or different (p > 0.05). We conclude that, in follow-up brain MRI, automatic positioning should be used, because manual positioning decreases the reproducibility of the scan prescription even if the same operator performs the second scan.
在后续的脑部磁共振成像(MRI)中,扫描方案的精确可重复性非常重要,这样可以防止对临床关注的体积变化估计过高或过低。(扫描方案定义为头部相对于三维MRI矩阵扫描平面的位置和方向。)在本研究中,计算了手动定位和自动定位情况下原始扫描与第二次扫描之间的配准误差。这些计算针对一名反复扫描的健康志愿者进行,并且在一项回顾性研究中,针对225名进行了原始扫描和至少一次后续扫描的患者进行。还研究了两次扫描由同一名或不同扫描操作员操作的影响。所有成像均使用市售的1.5特斯拉MRI系统和六元素头部阵列线圈。通过使用傅里叶相位相关方法将原始扫描图像与后续扫描图像配准来确定扫描方案的可重复性。我们的结果表明:(1)自动定位的可重复性优于手动定位(p < 0.05),并且(2)操作员相同或不同时结果无显著差异(p > 0.05)。我们得出结论,在后续脑部MRI中应使用自动定位,因为即使同一名操作员进行第二次扫描,手动定位也会降低扫描方案的可重复性。