The present study has attempted to downscale a mass spectrometer in order to make it portable and enable onsite analysis with it. The development of a small mass spectrometer required the use of a compact pump whose displacement was small, decreasing the sensitivity of that spectrometer. To get high sensitivity with a small mass spectrometer, we have integrated novel techniques: a highly sensitive ionization source and efficient extraction of sample vapor. The low-pressure dielectric barrier discharge ionization (LP-DBDI) source made it possible to increase the conductance between the source and the mass analyzer, compared with ambient ionization sources, enhancing the efficiency of the ion transfer from the ionization source to the mass analyzer. We have also developed a vacuumed headspace method efficiently transporting the sample vapor to the ionization source. The sensitivity was further enhanced by also using a discontinuous sample gas introduction technique. A prototype portable mass spectrometer using those novel techniques was found to be sensitive enough to detect 0.1 ppm methamphetamine, 1 ppm amphetamine, 1 ppm 3,4-methylenedioxymethamphetamine, and 10 ppm cocaine in liquid.
本研究试图缩小质谱仪的尺寸,使其便携并能够现场进行分析。小型质谱仪的开发需要使用排量小的紧凑型泵,这会降低该质谱仪的灵敏度。为了在小型质谱仪上获得高灵敏度,我们集成了新颖的技术:高灵敏度的离子源和高效提取样品蒸汽。与环境电离源相比,低压介电阻挡放电电离(LP-DBDI)源使源和质量分析器之间的电导率增加,提高了离子从电离源转移到质量分析器的效率。我们还开发了一种抽空顶空方法,可有效地将样品蒸汽输送到电离源。通过还使用不连续的样品气体引入技术,进一步提高了灵敏度。事实证明,使用这些新技术的原型便携式质谱仪足以灵敏地检测到液体中的 0.1 ppm 甲基苯丙胺、1 ppm 苯丙胺、1 ppm 3,4-亚甲二氧基甲基苯丙胺和 10 ppm 可卡因。