Department of Radiation Oncology, Samuel Oschin Comprehensive Cancer Institute, Cedars-Sinai Medical Center, Los Angeles, California 90048, USA.
Radiat Res. 2013 Apr;179(4):416-21. doi: 10.1667/RR2999.2.
In this study, we investigated the accuracy of using off-line bioluminescence imaging (BLI) and tomography (BLT) to guide irradiation of small soft tissue targets on a small animal radiation research platform (SARRP) with on-board cone beam CT (CBCT) capability. A small glass bulb containing BL cells was implanted as a BL source in the abdomen of 11 mouse carcasses. Bioluminescence imaging and tomography were acquired for each carcass. Six carcasses were setup visually without immobilization and 5 were restrained in position with tape. All carcasses were setup in treatment position on the SARRP where the centroid position of the bulb on CBCT was taken as "truth". In the 2D visual setup, the carcass was setup by aligning the point of brightest luminescence with the vertical beam axis. In the CBCT assisted setup, the pose of the carcass on CBCT was aligned with that on the 2D BL image for setup. For both 2D setup methods, the offset of the bulb centroid on CBCT from the vertical beam axis was measured. In the BLT-CBCT fusion method, the 3D torso on BLT and CBCT was registered and the 3D offset of the respective source centroids was calculated. The setup results were independent of the carcass being immobilized or not due to the onset of rigor mortis. The 2D offset of the perceived BL source position from the CBCT bulb position was 2.3 mm ± 1.3 mm. The 3D offset between BLT and CBCT was 1.5 mm ± 0.9 mm. Given the rigidity of the carcasses, the setup results represent the best that can be achieved with off-line 2D BLI and 3D BLT. The setup uncertainty would require the use of undesirably large margin of 4-5 mm. The results compel the implementation of on-board BLT capability on the SARRP to eliminate setup error and to improve BLT accuracy.
在这项研究中,我们调查了使用离线生物发光成像(BLI)和断层扫描(BLT)来指导具有车载锥形束 CT(CBCT)功能的小动物辐射研究平台(SARRP)上的小软组织靶区照射的准确性。将一个装有 BL 细胞的小玻璃灯泡作为 BL 源植入 11 只老鼠尸体的腹部。对每只尸体进行生物发光成像和断层扫描。六只尸体没有固定就进行了设置,五只则用胶带固定。所有尸体都在 SARRP 上的治疗位置进行设置,其中灯泡在 CBCT 上的质心位置被视为“真实”。在 2D 视觉设置中,通过将最亮发光点与垂直光束轴对齐来设置尸体。在 CBCT 辅助设置中,通过将 CBCT 上的尸体姿势与 2D BL 图像上的姿势对齐来设置。对于这两种 2D 设置方法,都测量了灯泡质心在 CBCT 上与垂直光束轴的偏移。在 BLT-CBCT 融合方法中,对 BLT 和 CBCT 上的 3D 躯干进行了注册,并计算了各自源质心的 3D 偏移。由于尸体开始僵硬,因此设置结果与尸体是否固定无关。感知的 BL 源位置与 CBCT 灯泡位置之间的 2D 偏移为 2.3mm±1.3mm。BLT 和 CBCT 之间的 3D 偏移为 1.5mm±0.9mm。考虑到尸体的刚性,该设置结果代表了离线 2D BLI 和 3D BLT 所能达到的最佳效果。设置不确定性将需要使用 4-5mm 的过大余量。结果迫使在 SARRP 上实施车载 BLT 能力,以消除设置误差并提高 BLT 精度。