尸体脊柱中双螺纹椎弓根螺钉即刻和疲劳强度的特点。

Characteristics of immediate and fatigue strength of a dual-threaded pedicle screw in cadaveric spines.

机构信息

Division of Neurological Surgery, Barrow Neurological Institute, St. Joseph's Hospital and Medical Center, 350 W. Thomas Rd, Phoenix, AZ 85013, USA.

出版信息

Spine J. 2013 Aug;13(8):947-56. doi: 10.1016/j.spinee.2013.03.010. Epub 2013 Apr 17.

Abstract

BACKGROUND CONTEXT

Novel dual-threaded screws are configured with overlapping (doubled) threads only in the proximal shaft to improve proximal cortical fixation.

PURPOSE

Tests were run to determine whether dual-threaded pedicle screws improve pullout resistance and increase fatigue endurance compared with standard pedicle screws.

STUDY DESIGN/SETTING: In vitro strength and fatigue tests were performed in human cadaveric vertebrae and in polyurethane foam test blocks.

PATIENT SAMPLE

Seventeen cadaveric lumbar vertebrae (14 pedicles) and 40 test sites in foam blocks were tested.

OUTCOME MEASURES

Measures for comparison between standard and dual-threaded screws were bone mineral density (BMD), screw insertion torque, ultimate pullout force, peak load at cyclic failure, and pedicular side of first cyclic failure.

METHODS

For each vertebral sample, dual-threaded screws were inserted in one pedicle and single-threaded screws were inserted in the opposite pedicle while recording insertion torque. In seven vertebrae, axial pullout tests were performed. In 10 vertebrae, orthogonal loads were cycled at increasing peak values until toggle exceeded threshold for failure. Insertion torque and pullout force were also recorded for screws placed in foam blocks representing healthy or osteoporotic bone porosity.

RESULTS

In bone, screw insertion torque was 183% greater with dual-threaded than with standard screws (p<.001). Standard screws pulled out at 93% of the force required to pull out dual-threaded screws (p=.42). Of 10 screws, five reached toggle failure first on the standard screw side, two screws failed first on the dual-threaded side, and three screws failed on both sides during the same round of cycling. In the high-porosity foam, screw insertion torque was 60% greater with the dual-threaded screw than with the standard screw (p=.005), but 14% less with the low-porosity foam (p=.07). Pullout force was 19% less with the dual-threaded screw than with the standard screw in the high-porosity foam (p=.115), but 6% greater with the dual-threaded screw in the low-porosity foam (p=.156).

CONCLUSIONS

Although dual-threaded screws required higher insertion torque than standard screws in bone and low density foam, dual-threaded and standard pedicle screws exhibited equivalent axial pullout and cyclic fatigue endurance. Unlike single-threaded screws, the mechanical performance of dual-threaded screws in bone was relatively independent of BMD. In foam, the mechanical performance of both types of screws was highly dependent on porosity.

摘要

背景

新型双螺纹螺钉仅在近侧轴上配置重叠(加倍)的螺纹,以改善近侧皮质固定。

目的

测试旨在确定与标准椎弓根螺钉相比,双螺纹椎弓根螺钉是否能提高抗拔出阻力并增加疲劳耐力。

研究设计/设置:在人体尸体椎体和聚氨酯泡沫测试块中进行体外强度和疲劳测试。

患者样本

17 个尸体腰椎(14 个椎弓根)和 40 个泡沫块测试部位。

测量结果

比较标准和双螺纹螺钉的测量值包括骨矿物质密度(BMD)、螺钉插入扭矩、最大拔出力、循环失效时的峰值载荷和第一次循环失效时的椎弓根侧。

方法

对于每个椎体样本,将双螺纹螺钉插入一个椎弓根,将单螺纹螺钉插入对侧椎弓根,并记录插入扭矩。在 7 个椎体中进行轴向拔出测试。在 10 个椎体中,以增加的峰值循环施加正交载荷,直到切换超过失败阈值。还记录了放置在代表健康或骨质疏松骨孔隙率的泡沫块中的螺钉的插入扭矩和拔出力。

结果

在骨中,双螺纹螺钉的插入扭矩比标准螺钉高 183%(p<.001)。标准螺钉拔出的力仅为双螺纹螺钉拔出所需力的 93%(p=.42)。在 10 个螺钉中,有 5 个首先在标准螺钉侧达到切换失效,2 个螺钉首先在双螺纹螺钉侧失效,3 个螺钉在同一循环中同时在两侧失效。在高孔隙率泡沫中,双螺纹螺钉的插入扭矩比标准螺钉高 60%(p=.005),但在低孔隙率泡沫中低 14%(p=.07)。在高孔隙率泡沫中,双螺纹螺钉的拔出力比标准螺钉低 19%(p=.115),但在低孔隙率泡沫中,双螺纹螺钉的拔出力比标准螺钉高 6%(p=.156)。

结论

尽管双螺纹螺钉在骨和低密度泡沫中比标准螺钉需要更高的插入扭矩,但双螺纹和标准椎弓根螺钉在轴向拔出和循环疲劳耐力方面表现出等效性。与单螺纹螺钉不同,双螺纹螺钉在骨中的机械性能相对独立于 BMD。在泡沫中,两种类型螺钉的机械性能都高度依赖于孔隙率。

文献AI研究员

20分钟写一篇综述,助力文献阅读效率提升50倍

立即体验

用中文搜PubMed

大模型驱动的PubMed中文搜索引擎

马上搜索