Amiss T J, Tyczkowska K L, Aucoin D P
Department of Anatomy, College of Veterinary Medicine, North Carolina State University, Raleigh 27606.
J Chromatogr. 1990 Apr 6;526(2):375-82. doi: 10.1016/s0378-4347(00)82521-9.
A liquid chromatographic method with fluorimetric detection was developed to measure taurine (2-aminoethanesulfonic acid) in feline plasma and whole blood. Plasma or lysed whole blood was diluted with a mixture of acetonitrile-methanol-triethylamine-water (25:22:3:50, v/v), filtered through a 10,000 dalton exclusion filter and derivatized with dansyl chloride for 30 min at room temperature. Dansyl taurine was separated from other compounds by reversed-phase liquid chromatography using an octadecyl column and a methanol-acetic acid-triethylamine (30:0.5:0.025, v/v) aqueous mobile phase. The effluent was monitored fluorimetrically at an excitation wavelength of 329 nm and an emission wavelength of 530 nm. The presence of mono-dansylated taurine in feline plasma was confirmed by thermospray mass spectrometry. The limit of detection was 16 nmol/ml and the detector response was linear from 40 to 4000 nmol/ml taurine.
建立了一种带有荧光检测的液相色谱法来测定猫血浆和全血中的牛磺酸(2-氨基乙磺酸)。血浆或裂解的全血用乙腈-甲醇-三乙胺-水(25:22:3:50,v/v)的混合物稀释,通过10,000道尔顿的排阻滤器过滤,并在室温下用丹磺酰氯衍生化30分钟。使用十八烷基柱和甲醇-乙酸-三乙胺(30:0.5:0.025,v/v)的水性流动相,通过反相液相色谱法将丹磺酰牛磺酸与其他化合物分离。在激发波长329nm和发射波长530nm下对流出物进行荧光监测。通过热喷雾质谱法确认了猫血浆中存在单丹磺酰化牛磺酸。检测限为16nmol/ml,检测器响应在40至4000nmol/ml牛磺酸范围内呈线性。