Division of Radiology, Department of Medicine and Health Sciences, Faculty of Health Sciences, Linköping University, Östergötland County Council, University Hospital, Linköping, Sweden; Center for Medical Image Science and Visualization (CMIV), Linköping University, Linköping, Sweden; Sectra Imtec AB, Linköping, Sweden.
Division of Radiology, Department of Clinical Science, Intervention and Technology, Karolinska Institutet, Karolinska University Hospital, Stockholm, Sweden.
J Clin Densitom. 2014 Jan-Mar;17(1):104-8. doi: 10.1016/j.jocd.2013.04.002. Epub 2013 May 7.
During the past 15 yr, digital X-ray radiogrammetry (DXR) has been used to measure metacarpal bone mineral density (BMD). BMD is often measured in existing cohorts where X-ray images were not acquired in accordance with the DXR imaging protocol (DIP). The purpose of the present study was to analyze how deviations from DIP in historical radiographs may affect the reproducibility of DXR-BMD measurements. Cadaver hand phantoms were used to conduct repeat measurements of deviations from DIP with respect to voltage, exposure, lateral displacement, supination, combination of lateral displacement and supination or rotation, extension of the wrist, and edge enhancement. Direct digital radiography (Aristos; Siemens Healthcare, Erlangen, Germany) was used for image acquisition and dxr-online (Sectra, Linköping, Sweden) for DXR-BMD measurements. The influence of the tested deviations from DIP ranged from 0 to 32.5 mg/cm(2) (0-6.8%). On repetition with the same specimen, none of the deviations resulted in a within-specimen reproducibility error greater than 2 mg/cm(2) (0.4%, equivalent to a T-score of 0.042). Among the tested deviations, all except tube voltage had a magnitude greater than the normal measurement noise for the technique and must therefore be considered when planning a study based on historical images.
在过去的 15 年中,数字 X 射线射线照相术(DXR)已被用于测量掌骨骨密度(BMD)。在现有的队列中,通常会测量 BMD,其中 X 射线图像的获取不符合 DXR 成像协议(DIP)。本研究的目的是分析历史射线照片中偏离 DIP 如何影响 DXR-BMD 测量的可重复性。使用尸体手模型进行了对 DIP 的偏离的重复测量,涉及电压、曝光、侧向位移、旋前、侧向位移和旋前的组合或旋转、腕部伸展以及边缘增强。直接数字射线照相(Aristos;西门子医疗,埃朗根,德国)用于图像采集,dxr-online(Sectra,林雪平,瑞典)用于 DXR-BMD 测量。从 0 到 32.5 mg/cm²(0-6.8%)的 DIP 偏离范围对测试结果有影响。对于同一样本进行重复测量,没有任何偏离导致体内重复性误差大于 2 mg/cm²(0.4%,相当于 T 评分 0.042)。在测试的偏离中,除了管电压之外,所有偏离都大于该技术的正常测量噪声,因此在基于历史图像的研究计划中必须考虑这些偏离。