Department of Cardiology, Gifu Prefectural General Medical Center, Gifu, Japan.
J Cardiol. 2013 Sep;62(3):188-94. doi: 10.1016/j.jjcc.2013.03.011. Epub 2013 May 11.
The aim of this study was to validate the accuracy of three-dimensional (3D) speckle tracking echocardiography (STE) and two-dimensional (2D)-STE for the assessment of left atrial (LA) volume and function by comparison with 3D-computed tomography (CT) performed on the same day as STE.
LA phasic volume and emptying function (EF) were measured in 28 patients with paroxysmal atrial fibrillation undergoing catheter ablation (62±11 years old) using both 3D-STE and 2D-STE during sinus rhythm. LA phasic volume and function measured by 3D-STE and 2D-STE were validated using 3D-CT as a gold standard.
The intraobserver correlation coefficient and variability in maximum LA volume assessed by 3D-STE were 0.99 and 1.4±6.0%, respectively. The interobserver correlation coefficient and variability in maximum LA volume assessed by 3D-STE were 0.99 and 0.2±4.5%, respectively. There were strong correlations between LA phasic volume measured by 3D-CT and those measured by 3D-STE (r=0.98, p<0.001). There were correlations between LA phasic function measured by 3D-CT and those measured by 3D-STE (r=0.85-0.88, p<0.001). There was a better agreement between 3D-CT and 3D-STE in the assessment of LA phasic volumes and function than between 3D-CT and 2D-STE in apical 2- and 4-chamber view.
3D-STE allows more accurate measurement of LA volume and function than 2D-STE and has high reproducibility.
本研究旨在通过与同一天进行的三维(3D)计算机断层扫描(CT)比较,验证三维斑点追踪超声心动图(STE)和二维(2D)-STE 评估左心房(LA)容积和功能的准确性。
在 28 例阵发性心房颤动行导管消融术的患者中(62±11 岁),在窦性心律下使用 3D-STE 和 2D-STE 测量 LA 时相容积和排空功能(EF)。使用 3D-CT 作为金标准验证通过 3D-STE 和 2D-STE 测量的 LA 时相容积和功能。
3D-STE 评估最大 LA 容积的观察者内相关系数和变异性分别为 0.99 和 1.4±6.0%。3D-STE 评估最大 LA 容积的观察者间相关系数和变异性分别为 0.99 和 0.2±4.5%。3D-CT 测量的 LA 时相容积与 3D-STE 测量的容积之间存在很强的相关性(r=0.98,p<0.001)。3D-CT 测量的 LA 时相功能与 3D-STE 测量的功能之间存在相关性(r=0.85-0.88,p<0.001)。与 3D-CT 和 2D-STE 在心尖 2 腔和 4 腔视图相比,3D-CT 和 3D-STE 在评估 LA 时相容积和功能方面具有更好的一致性。
3D-STE 比 2D-STE 更能准确测量 LA 容积和功能,且具有更高的可重复性。