Center for Bioanalysis, Division of Metrology for Quality of Life, Korea Research Institute of Standards and Science, 267 Gajeong-ro, Yuseong-gu, Daejeon, 305-340, Republic of Korea.
Anal Bioanal Chem. 2013 Oct;405(25):8063-72. doi: 10.1007/s00216-013-6999-6. Epub 2013 May 15.
We describe a capillary electrophoresis-mass spectrometry (CE-MS) method for newborn screening of a representative amino acid metabolic disease, namely, phenylketonuria (PKU). Underivatized phenylalanine and tyrosine in a dried blood spot (DBS) were simultaneously determined by CE-MS equipped with an ionophore membrane-packed sheathless electrospray ionization interface, which was developed by our group. The method was optimized for rapid determination of the underivatized amino acids, phenylalanine and tyrosine extracted from a DBS. Under the optimized conditions, the limit of detection of phenylalanine and tyrosine (signal-to-noise ratio, 3) was 0.03 and 0.07 mg/L in DBS, respectively, with a CE run time of less than 3 min. For repeated runs of a sample, coefficients of variation (CVs) for migration time were less than 3.7%, whereas CVs for the area ratio under the curve were 2.1 and 2.9% for 20 consecutive runs of 49.5 mg/kg Phe and 36.2 mg/kg Tyr, respectively. However, the relative standard deviations of intra- and interday assays for DBS samples were <6.2 and <5.8%, respectively, which were substantially due to sample extraction from DBS. The analytical method was applied to real clinical samples of Korean neonates, and results were compared with those of conventional methods for PKU diagnosis, which required reference analytical methods such as isotope dilution CE-MS or high-performance liquid chromatography-mass spectrometry for quality assurance of the conventional kit-based assays. The distinct advantages of high sensitivity and extremely low sample volume, as well as a simple, easy, and economic sample pretreatment, were demonstrated for the proposed method.
我们描述了一种毛细管电泳-质谱(CE-MS)方法,用于代表氨基酸代谢疾病的新生儿筛查,即苯丙酮尿症(PKU)。通过我们小组开发的带有载流膜无鞘电喷雾接口的 CE-MS,同时测定了干血斑(DBS)中的未衍生苯丙氨酸和酪氨酸。该方法针对从 DBS 中提取的未衍生氨基酸苯丙氨酸和酪氨酸的快速测定进行了优化。在优化条件下,DBS 中苯丙氨酸和酪氨酸(信噪比为 3)的检测限分别为 0.03 和 0.07 mg/L,CE 运行时间小于 3 分钟。对于样品的重复运行,迁移时间的变异系数(CV)小于 3.7%,而对于 20 次连续运行 49.5 mg/kg Phe 和 36.2 mg/kg Tyr 的曲线下面积比的 CV 分别为 2.1%和 2.9%。然而,DBS 样品的日内和日间测定的相对标准偏差分别小于 6.2%和 5.8%,这主要归因于从 DBS 中提取样品。该分析方法应用于韩国新生儿的实际临床样本,并将结果与传统 PKU 诊断方法进行比较,这需要参考分析方法,如同位素稀释 CE-MS 或高效液相色谱-质谱,以保证传统试剂盒分析的质量。该方法具有灵敏度高、样品体积极小、样品预处理简单、易于操作和经济等优点。