Suppr超能文献

高除颤阈值管理中除颤向量的概念

Concept of defibrillation vector in the management of high defibrillation threshold.

作者信息

Hayes Kevin, Deshmukh Abhishek, Pant Sadip, Tobler Gareth, Paydak Hakan

机构信息

Kevin Hayes, Abhishek Deshmukh, Hakan Paydak, Department of Cardiology, University of Arkansas for Medical Sciences, Little Rock, AR 72205, United States.

出版信息

World J Cardiol. 2013 Apr 26;5(4):106-8. doi: 10.4330/wjc.v5.i4.106.

Abstract

We present a case where defibrillation threshold was dangerously elevated to the point that the patient had no safety margin, and his implantable cardioverter-defibrillator generator was discovered to have migrated. Generator migration reduces the distance between the can and the coil, effectively creating a smaller bipolar current and sparing the left ventricle from the current needed for defibrillation. This case underscores the importance of securing the generator in place, as this patient would have been spared multiple shocks and an invasive medical procedure had his generator been better secured.

摘要

我们报告了一例除颤阈值危险地升高至患者没有安全裕度的病例,并且发现其植入式心脏复律除颤器发生器发生了移位。发生器移位缩短了罐体与线圈之间的距离,实际上产生了较小的双极电流,使左心室免受除颤所需电流的影响。该病例强调了将发生器固定到位的重要性,因为如果患者的发生器固定得更好,他本可避免多次电击和侵入性医疗程序。

相似文献

1
Concept of defibrillation vector in the management of high defibrillation threshold.
World J Cardiol. 2013 Apr 26;5(4):106-8. doi: 10.4330/wjc.v5.i4.106.
6
Implantation of a unipolar cardioverter/defibrillator system under local anaesthesia.
Eur Heart J. 1996 May;17(5):764-8. doi: 10.1093/oxfordjournals.eurheartj.a014944.
7
Prospective evaluation of the effect of biphasic waveform defibrillation on ventricular pacing thresholds.
J Cardiovasc Electrophysiol. 1997 May;8(5):485-95. doi: 10.1111/j.1540-8167.1997.tb00816.x.
9
Implantable cardioverter defibrillator lead placement in the middle cardiac vein after tricuspid valve surgery.
Europace. 2012 Jun;14(6):853-8. doi: 10.1093/europace/eus013. Epub 2012 Feb 14.

引用本文的文献

本文引用的文献

1
Low doses of intravenous epinephrine for refractory sustained monomorphic ventricular tachycardia.
World J Cardiol. 2012 Oct 26;4(10):296-301. doi: 10.4330/wjc.v4.i10.296.
2
The entirely subcutaneous implantable cardioverter-defibrillator: initial clinical experience in a large Dutch cohort.
J Am Coll Cardiol. 2012 Nov 6;60(19):1933-9. doi: 10.1016/j.jacc.2012.06.053. Epub 2012 Oct 10.
3
A randomized study of defibrillator lead implantations in the right ventricular mid-septum versus the apex: the SEPTAL study.
J Cardiovasc Electrophysiol. 2012 Aug;23(8):853-60. doi: 10.1111/j.1540-8167.2012.02311.x. Epub 2012 Mar 27.
4
High defibrillation threshold: the science, signs and solutions.
Indian Pacing Electrophysiol J. 2010 Jan 7;10(1):21-39.
5
High defibrillation threshold in patients with implantable defibrillator: how effective is the subcutaneous finger lead?
Eur J Cardiothorac Surg. 2009 Mar;35(3):489-92. doi: 10.1016/j.ejcts.2008.10.021. Epub 2009 Jan 14.
6
How to manage the patient with a high defibrillation threshold.
Heart Rhythm. 2006 Apr;3(4):492-5. doi: 10.1016/j.hrthm.2005.12.023.

文献AI研究员

20分钟写一篇综述,助力文献阅读效率提升50倍。

立即体验

用中文搜PubMed

大模型驱动的PubMed中文搜索引擎

马上搜索

文档翻译

学术文献翻译模型,支持多种主流文档格式。

立即体验