Department of Radiology, College of Medicine, Mayo Clinic, 200 First St SW, Rochester, MN 55905, USA.
AJR Am J Roentgenol. 2013 Jun;200(6):W621-7. doi: 10.2214/AJR.12.9413.
The purpose of this study is to evaluate radiation dose reduction strategies in perfusion CT by using a biologic phantom.
A formalin-preserved porcine liver was submerged in a 32-cm-wide acrylic phantom filled with water. The portal vein was connected to a continuous flow pump. The phantom was scanned with a perfusion CT protocol using 80 kVp and 400 mAs, every 1 second, for 50 seconds. This was repeated using 100 and 20 mAs. It was also repeated again using 400 mAs to assess reproducibility. A sparser scan frequency was simulated retrospectively. Blood flow was determined for each dataset using the maximum slope and deconvolution methods.
Measurements of the mean blood flow values in identical regions of interest had a percent difference of 7% for repeated perfusion CT protocols using the same settings regardless of perfusion model used. The 100 mAs scans agreed with 400 mAs scans, with percent differences of 21% and 31% for the maximum slope and deconvolution methods, respectively. At a simulated frequency of one scan every 4 seconds, blood flow values differed up to 17% and 60% from the reference scan for the maximum slope and deconvolution methods, respectively. At 20 mAs and one scan every 1 second, or 400 mAs and a simulated frequency of one scan every 8 seconds, both computation methods failed to provide accurate blood flow estimates.
The biologic phantom showed reproducible measurements that can help in optimizing perfusion CT protocols by determining both the acquisition parameters that affect radiation dose and the accuracy of estimates from different perfusion models.
本研究旨在通过生物体模评估灌注 CT 中的辐射剂量降低策略。
福尔马林固定的猪肝浸泡在充满水的 32cm 宽的丙烯酸体模中。门静脉与连续流动泵相连。使用灌注 CT 协议以 80kVp 和 400mAs 扫描体模,每秒 1 次,持续 50 秒。重复使用 100 和 20mAs 进行重复扫描。再次重复使用 400mAs 以评估可重复性。回顾性模拟更稀疏的扫描频率。使用最大斜率和去卷积方法为每个数据集确定血流。
无论使用何种灌注模型,相同设置的重复灌注 CT 协议测量相同 ROI 的平均血流值之间的百分比差异为 7%。100mAs 扫描与 400mAs 扫描一致,最大斜率和去卷积方法的百分比差异分别为 21%和 31%。在模拟的每 4 秒扫描一次的频率下,最大斜率和去卷积方法的血流值与参考扫描相差高达 17%和 60%。在 20mAs 和每 1 秒扫描一次,或 400mAs 和模拟的每 8 秒扫描一次的频率下,两种计算方法都无法提供准确的血流估计值。
生物体模显示出可重复的测量结果,通过确定影响辐射剂量的采集参数以及不同灌注模型的估计准确性,有助于优化灌注 CT 方案。