Caobelli F, Pizzocaro C, Paghera B, Guerra U P
Dr. Federico Caobelli, Department of Nuclear Medicine, Fondazione Poliambulanza, Via Bissolati 57, 25100 Brescia,Italy, Tel./Fax +39/030/351 81 04, E-mail:
Nuklearmedizin. 2013;52(5):178-85. doi: 10.3413/Nukmed-0570-13-03. Epub 2013 May 24.
Quantification of myocardial perfusion scintigraphy is often performed to assist physicians in detecting coronary artery disease (CAD). Modern software and hardware packages provide improvements able to shorten scan time and/or reduce administered activity, without compromising image quality in radionuclide myocardial perfusion imaging (MPI). Recently, multifocal collimators were introduced with dedicated reconstruction software, named IQ-SPECT, able to shorten considerably scan time. The aim of our study was to compare this new protocol to the already validated standard ones.
PATIENTS, METHODS: We enrolled 43 patients with suspected or diagnosed CAD. All patients underwent a two-days protocol radionuclide myocardial perfusion scan at rest and after a standard stress test (exercise or dipyridamole) after administering 99mTc-tetrofosmin. Images were acquired on a 2-head gamma camera and reconstructed with attenuation correction. All the images were scored using a 17-segments model by three experienced physicians, blind to clinical data and to acquisition and processing modality.
RESULTS, CONCLUSION: No significant differences were recorded in perfusion scores on paired t-test and Wilcoxon among the full-time images reconstructed with standard protocol or IQ-SPECT, both overall on a 17-segments evaluation and when considering different territories of distribution. MPI with IQ-SPECT protocol can be acquired at about a quarter scan time without disagreement compared to full time scan acquisition performed with standard protocols.
心肌灌注闪烁扫描的定量分析常被用于辅助医生检测冠状动脉疾病(CAD)。现代软件和硬件包带来了改进,能够缩短扫描时间和/或减少给药剂量,同时不影响放射性核素心肌灌注成像(MPI)的图像质量。最近,引入了多焦点准直器及名为IQ-SPECT的专用重建软件,能够大幅缩短扫描时间。我们研究的目的是将这种新方案与已得到验证的标准方案进行比较。
患者、方法:我们纳入了43例疑似或确诊CAD的患者。所有患者在静息状态下以及在给予99m锝-替曲膦后进行标准负荷试验(运动或双嘧达莫)后,接受为期两天的放射性核素心肌灌注扫描。图像在双头γ相机上采集,并进行衰减校正重建。所有图像由三位经验丰富的医生使用17节段模型进行评分,医生对临床数据以及采集和处理方式不知情。
结果、结论:在使用标准方案或IQ-SPECT重建的全时图像上,无论是在17节段评估总体上还是考虑不同分布区域时,配对t检验和Wilcoxon检验的灌注评分均未记录到显著差异。与使用标准方案进行的全时扫描采集相比,采用IQ-SPECT方案的MPI扫描时间可缩短约四分之三,且无分歧。