J Nucl Cardiol. 2018 Aug;25(4):1328-1337. doi: 10.1007/s12350-016-0770-2. Epub 2017 Jan 3.
Image acquisition by short-time single-photon emission-computed tomography (SPECT) has been made feasible by IQ·SPECT. The aim of this study was to generate normal databases (NDBs) of thallium-201 (Tl) myocardial perfusion imaging for IQ·SPECT, and characterize myocardial perfusion distribution.
We retrospectively enrolled 159 patients with a low likelihood of cardiac diseases from four hospitals in Japan. All patients underwent short-time Tl myocardial perfusion IQ·SPECT with or without attenuation and scatter correction (ACSC) in either supine or prone position. The mean myocardial counts were calculated using 17-segment polar maps. Three NDBs were derived from supine and prone images as well as supine images with ACSC. Differences between the supine and prone positions were observed in the uncorrected sex-segregated NDBs in the mid-inferolateral counts (p ≤ 0.016 for males and p ≤ 0.002 for females). Differences between IQ·SPECT and conventional SPECT were also observed in the mid-anterior, inferolateral, and apical lateral counts (p ≤ 0.009 for males and p ≤ 0.003 for females). Apical low counts attributed to myocardial thinning were observed in the apical anterior and apex segments in the supine IQ·SPECT NDB with ACSC.
There were significant differences between uncorrected supine and prone NDBs, between uncorrected supine NDB and supine NDB with ACSC, and between uncorrected supine NDB and conventional SPECT NDB. Understanding the pattern of normal distribution in IQ-SPECT short-time acquisitions with and without ACSC will be helpful for interpretation of imaging findings in patients with coronary artery disease (CAD) or low likelihood of CAD and the NDBs will aid in quantitative analysis.
通过 IQ·SPECT 实现了短时间单光子发射计算机断层扫描(SPECT)的图像采集。本研究旨在为 IQ·SPECT 生成铊-201(Tl)心肌灌注成像的正常数据库(NDB),并描述心肌灌注分布特征。
我们回顾性纳入了来自日本四家医院的 159 例低心脏疾病可能性患者。所有患者均行短时间 Tl 心肌灌注 IQ·SPECT 检查,包括仰卧位或俯卧位,是否进行衰减和散射校正(ACSC)。采用 17 节段极坐标图计算平均心肌计数。根据仰卧位和俯卧位图像以及仰卧位 ACSC 图像得出 3 个 NDB。在未经校正的男女分组 NDB 中,仰卧位和俯卧位之间存在中侧下计数的差异(男性 p≤0.016,女性 p≤0.002)。在男性和女性中,IQ·SPECT 和常规 SPECT 之间也存在中前、侧下和侧尖计数的差异(p≤0.009 和 p≤0.003)。在仰卧位 ACSC IQ·SPECT NDB 中,前尖和心尖段存在心尖变薄导致的低心尖计数。
未校正仰卧位和俯卧位 NDB 之间、未校正仰卧位 NDB 和仰卧位 ACSC NDB 之间以及未校正仰卧位 NDB 和常规 SPECT NDB 之间存在显著差异。了解 ACSC 存在和不存在时 IQ-SPECT 短时间采集的正常分布模式将有助于解释冠心病(CAD)或低 CAD 可能性患者的成像发现,并且 NDB 将有助于定量分析。