使用动态体模评估心肌灌注成像中的射血分数:IQ-SPECT与低能高分辨率准直器(LEHR)的比较

Ejection fraction in myocardial perfusion imaging assessed with a dynamic phantom: comparison between IQ-SPECT and LEHR.

作者信息

Hippeläinen Eero, Mäkelä Teemu, Kaasalainen Touko, Kaleva Erna

机构信息

HUS Medical Imaging Center, Clinical Physiology and Nuclear Medicine, Helsinki University Hospital, University of Helsinki, POB 340, FI-00029, HUS, Helsinki, Finland.

Department of Physics, University of Helsinki, Helsinki, Finland.

出版信息

EJNMMI Phys. 2017 Dec;4(1):20. doi: 10.1186/s40658-017-0187-2. Epub 2017 Jul 19.

Abstract

BACKGROUND

Developments in single photon emission tomography instrumentation and reconstruction methods present a potential for decreasing acquisition times. One of such recent options for myocardial perfusion imaging (MPI) is IQ-SPECT. This study was motivated by the inconsistency in the reported ejection fraction (EF) and left ventricular (LV) volume results between IQ-SPECT and more conventional low-energy high-resolution (LEHR) collimation protocols. IQ-SPECT and LEHR quantitative results were compared while the equivalent number of iterations (EI) was varied. The end-diastolic (EDV) and end-systolic volumes (ESV) and the derived EF values were investigated. A dynamic heart phantom was used to produce repeatable ESVs, EDVs and EFs. Phantom performance was verified by comparing the set EF values to those measured from a gated multi-slice X-ray computed tomography (CT) scan (EF). The phantom with an EF setting of 45, 55, 65 and 70% was imaged with both IQ-SPECT and LEHR protocols. The data were reconstructed with different EI, and two commonly used clinical myocardium delineation software were used to evaluate the LV volumes.

RESULTS

The CT verification showed that the phantom EF settings were repeatable and accurate with the EF being within 1% point from the manufacture's nominal value. Depending on EI both MPI protocols can be made to produce correct EF estimates, but IQ-SPECT protocol produced on average 41 and 42% smaller EDV and ESV when compared to the phantom's volumes, while LEHR protocol underestimated volumes by 24 and 21%, respectively. The volume results were largely similar between the delineation methods used.

CONCLUSIONS

The reconstruction parameters can greatly affect the volume estimates obtained from perfusion studies. IQ-SPECT produces systematically smaller LV volumes than the conventional LEHR MPI protocol. The volume estimates are also software dependent.

摘要

背景

单光子发射断层扫描仪器和重建方法的发展为缩短采集时间带来了潜力。心肌灌注成像(MPI)的一种最新选择是IQ-SPECT。本研究的动机是IQ-SPECT与更传统的低能高分辨率(LEHR)准直协议之间报告的射血分数(EF)和左心室(LV)容积结果不一致。在改变等效迭代次数(EI)的同时,比较了IQ-SPECT和LEHR的定量结果。研究了舒张末期(EDV)和收缩末期容积(ESV)以及导出的EF值。使用动态心脏模型来产生可重复的ESV、EDV和EF。通过将设定的EF值与从门控多层X射线计算机断层扫描(CT)扫描测量的EF值进行比较,验证了模型性能。使用IQ-SPECT和LEHR协议对EF设置为45%、55%、65%和70%的模型进行成像。数据用不同的EI进行重建,并使用两种常用的临床心肌描绘软件来评估LV容积。

结果

CT验证表明,模型EF设置是可重复且准确的,EF与制造商的标称值相差在1个百分点以内。根据EI,两种MPI协议都可以得出正确的EF估计值,但与模型容积相比,IQ-SPECT协议产生的EDV和ESV平均分别小41%和42%,而LEHR协议低估容积分别为24%和21%。所使用的描绘方法之间的容积结果在很大程度上相似。

结论

重建参数可极大地影响灌注研究获得的容积估计值。与传统的LEHR MPI协议相比,IQ-SPECT产生的LV容积系统性地更小。容积估计值也依赖于软件。