Guerra Jose M, Jorge Esther, Raga Silvia, Gálvez-Montón Carolina, Alonso-Martín Concepción, Rodríguez-Font Enrique, Cinca Juan, Viñolas Xavier
Department of Cardiology, Hospital de la Santa Creu i Sant Pau, Universitat Autònoma de Barcelona, Institut de Recerca HSCSP-IIB Sant Pau, Barcelona, Spain.
J Cardiovasc Electrophysiol. 2013 Oct;24(10):1157-62. doi: 10.1111/jce.12175. Epub 2013 May 29.
Open-irrigated radiofrequency ablation catheters with slight differences in tip architecture are widely used, although limited comparative data are available. The purpose of this study was to compare the lesion size and potential complications produced by commercially available open-irrigated catheters in an in vitro porcine heart model.
Six catheters were tested (Biosense Webster Thermocool, Boston Scientific Open irrigated, St. Jude CoolPath, St. Jude CoolPath Duo, Biosense Webster Thermocool SF, St. Jude Cool Flex) at 20 and 35 W power-control, under 2 different blood flows (0.1 and 0.5 m/s) and at 2 target durations (30 and 60 seconds). A total of 601 lesions were made in 26 in vitro preparations. The tip temperature profile showed significant differences between the catheters (P < 0.001) with the Thermocool SF registering the lowest. Only the surface diameter and the depth at maximum diameter of the lesion were influenced by the design of the ablation electrode. The lesion volume did not show significant differences between catheters for any power, application duration or blood flow condition. Char and pops occurred more often at 35 W with only slight differences between the catheters.
Tip design of the 6 different irrigated catheters does not affect the lesion total volume, although a slight difference in lesion geometry in terms of surface diameter and depth at maximum diameter is present. The catheters show a slight different in vitro safety profile.
尽管可用的比较数据有限,但尖端结构略有差异的开放式灌注射频消融导管仍被广泛使用。本研究的目的是在体外猪心脏模型中比较市售开放式灌注导管产生的病变大小和潜在并发症。
在20瓦和35瓦功率控制下,于2种不同血流速度(0.1米/秒和0.5米/秒)以及2种目标持续时间(30秒和60秒)条件下,对6种导管进行了测试(Biosense Webster Thermocool、Boston Scientific Open irrigated、St. Jude CoolPath、St. Jude CoolPath Duo、Biosense Webster Thermocool SF、St. Jude Cool Flex)。在26个体外样本中总共制造了601个病变。尖端温度曲线显示导管之间存在显著差异(P < 0.001),其中Thermocool SF记录的温度最低。只有病变的表面直径和最大直径处的深度受消融电极设计的影响。对于任何功率、应用持续时间或血流条件,导管之间的病变体积均未显示出显著差异。炭化和爆裂在35瓦时更常发生,导管之间只有细微差异。
6种不同灌注导管的尖端设计不影响病变总体积,尽管在病变几何形状方面,表面直径和最大直径处的深度存在细微差异。这些导管在体外安全性方面表现出细微差异。