J Vet Intern Med. 2013 Jul-Aug;27(4):990-5. doi: 10.1111/jvim.12115. Epub 2013 Jun 4.
Hospitalized alpacas are often hyperglycemic requiring frequent blood glucose testing.
To compare the performance of 4 brands of glucometers with a laboratory-based analyzer (LCA) over a range of glucose concentrations in alpacas.
Four healthy male alpacas.
A 2-treatment cross-over study was utilized. The alpacas were given 0.4 U/kg of regular insulin intravenously and then 500 mg/kg of dextrose intravenously with a 1 week washout period between treatments. Blood samples were collected from 10 minutes before until 6 hours after drug administration. Glucose concentrations were measured in whole blood and plasma samples on 4 glucometers, and serum glucose was measured on an LCA.
Glucometer performance varied depending on whether glucose concentrations were measured in plasma or whole blood. Based on error grid analysis, the Precision Xtra and One Touch Ultra 2 glucometers were clinically acceptable for testing whole blood samples, whereas the Accu-Chek Aviva and Nova StatStrip Xpress glucometers were clinically acceptable for testing plasma samples in comparison with serum glucose concentrations determined by the LCA. All glucometers had systematic and proportional biases that varied based on sample type.
Human-based glucometers in alpacas should be used cautiously, particularly at higher blood glucose concentrations. The blood sample type (plasma or whole blood) can alter meter performance when compared with serum glucose concentrations and potentially lead to errors in clinical decisions.
住院的羊驼通常会出现高血糖,需要频繁进行血糖检测。
比较 4 种血糖仪与实验室基础分析仪(LCA)在羊驼不同血糖浓度范围内的性能。
4 只健康雄性羊驼。
采用 2 种处理的交叉研究。羊驼静脉注射 0.4 U/kg 的常规胰岛素,然后静脉注射 500mg/kg 的葡萄糖,两种处理之间有 1 周的洗脱期。在给药前 10 分钟至给药后 6 小时采集血液样本。在 4 种血糖仪上测量全血和血浆样本中的葡萄糖浓度,在 LCA 上测量血清葡萄糖浓度。
血糖仪的性能取决于测量血浆或全血中的葡萄糖浓度。根据误差网格分析,Precision Xtra 和 One Touch Ultra 2 血糖仪在测试全血样本时具有临床可接受性,而 Accu-Chek Aviva 和 Nova StatStrip Xpress 血糖仪在测试血浆样本时与 LCA 测定的血清葡萄糖浓度相比具有临床可接受性。所有血糖仪都有系统和比例偏差,这些偏差取决于样本类型。
在羊驼中使用基于人体的血糖仪时应谨慎,尤其是在血糖浓度较高时。与血清葡萄糖浓度相比,血液样本类型(血浆或全血)会改变血糖仪的性能,并可能导致临床决策错误。