基于 click-evoked otoacoustic emission 的内侧橄榄耳蜗传出神经功能测试的可重复性。

Repeatability of click-evoked otoacoustic emission-based medial olivocochlear efferent assay.

机构信息

Institute of Sound and Vibration Research, Hearing & Balance Center, University of Southampton, Southampton, United Kingdom.

出版信息

Ear Hear. 2013 Nov-Dec;34(6):789-98. doi: 10.1097/AUD.0b013e3182944c04.

Abstract

OBJECTIVES

Otoacoustic emissions-based efferent assays are evolving to become a part of auditory diagnostics. The wide range of clinical applications, such as assessment of auditory neuropathy, auditory processing disorders, learning disability, monitoring success in auditory intervention and others illustrate the significance of this measurement. Defining the procedure's test-retest repeatability is of critical importance, to allow for distinction between measurement deviations and true physiological or pathological changes. The purpose of this study was to assess the repeatability of a click-evoked otoacoustic emission-based (CEOAE) test of the medial olivocochlear (MOC) reflex in normal-hearing (NH) adults.

DESIGN

Test-retest data were collected from 35 NH young adults in two distinct test sessions separated by 1 to 4 days. CEOAEs were recorded without and with contralateral acoustic stimulation (CAS; 35 dB SL). Three indices of the MOC reflex were computed: CAS-induced (a) absolute changes in CEOAE amplitude, (b) normalized changes in CEOAE amplitude, and (c) changes in CEOAE input-output functions. Repeatability of these indices was assessed by a three-layered approach, which consisted of Bland-Altman plots, coefficient of reliability (Cronbach's α), and analysis of variance.

RESULTS

Analyses indicated good repeatability of three CEOAE-based MOC reflex indices. A two-way analysis of variance of the indices demonstrated no significant difference between test and retest. Normalized index showed similar repeatability as other indices. CEOAE signal to noise ratio did not seem to vary between test sessions. Notably, CAS caused a decrease in CEOAE input-output functions slope in a majority of participants (n = 29).

CONCLUSIONS

The present study is the first to elucidate the intrasubject variability of absolute and normalized indices of the MOC inhibitory effect. Although the measurements were conducted under realistic conditions resembling the clinical setting, repeatability was generally good in NH adults. For MOC reflex test, the signal to noise ratio of 6 dB for recording CEOAEs seems to be a recommendable criterion when considering practicability and measurement quality in clinical conditions. The present findings exemplify the suitability of CEOAE-based MOC assay as a monitoring tool of medial efferent status over time. The data are intended to assist clinicians and scientists alike in the accurate interpretation of CAS-induced CEOAE changes in the test-retest situation.

摘要

目的

基于耳声发射的传出测试正在发展成为听觉诊断的一部分。这种测试具有广泛的临床应用,例如评估听觉神经病、听觉处理障碍、学习障碍、监测听觉干预的成功等,这说明了这种测量的重要性。定义该测试的重测重复性非常重要,以便能够区分测量偏差和真正的生理或病理变化。本研究的目的是评估正常听力(NH)成年人中基于click 诱发耳声发射的(CEOAE)中内侧橄榄耳蜗(MOC)反射测试的重复性。

设计

在两次不同的测试中,从 35 名 NH 年轻成年人中收集了测试-重测数据,两次测试之间间隔 1 至 4 天。在没有和有对侧声刺激(CAS;35dB SL)的情况下记录 CEOAEs。计算了 MOC 反射的三个指标:(a)CAS 引起的 CEOAE 幅度的绝对变化;(b)CEOAE 幅度的归一化变化;和(c)CEOAE 输入-输出函数的变化。通过三层方法评估这些指标的重复性,该方法包括 Bland-Altman 图、可靠性系数(Cronbach's α)和方差分析。

结果

分析表明,三种基于 CEOAE 的 MOC 反射指标的重复性良好。对这些指数的双向方差分析表明,测试和复测之间没有显著差异。归一化指数显示出与其他指数相似的重复性。在两次测试中,CEOAE 信号与噪声比似乎没有变化。值得注意的是,在大多数参与者(n = 29)中,CAS 导致 CEOAE 输入-输出函数斜率下降。

结论

本研究首次阐明了 MOC 抑制效应的绝对和归一化指数的个体内可变性。虽然测量是在类似于临床环境的现实条件下进行的,但 NH 成年人的重复性通常较好。对于 MOC 反射测试,在考虑临床条件下的实用性和测量质量时,记录 CEOAEs 的信噪比为 6dB 似乎是一个推荐的标准。本研究结果证明了基于 CEOAE 的 MOC 测定作为一种随时间监测内侧传出状态的监测工具的适用性。这些数据旨在帮助临床医生和科学家在测试-重测情况下准确解释 CAS 引起的 CEOAE 变化。