Vidas系统与两种近期用于孕妇和新生儿弓形虫病诊断及随访的全自动检测方法的比较。

Comparison of the Vidas system and two recent fully automated assays for diagnosis and follow-up of toxoplasmosis in pregnant women and newborns.

作者信息

Murat Jean-Benjamin, Dard Céline, Fricker Hidalgo Hélène, Dardé Marie-Laure, Brenier-Pinchart Marie-Pierre, Pelloux Hervé

机构信息

Laboratory of Parasitology and Mycology, Grenoble University Hospital, Grenoble, France.

出版信息

Clin Vaccine Immunol. 2013 Aug;20(8):1203-12. doi: 10.1128/CVI.00089-13. Epub 2013 Jun 5.

Abstract

Serological testing to detect toxoplasmosis is of major importance to avoid the possible effects of the disease in newborns. This study assessed anti-Toxoplasma IgG and IgM with the Vidas (bioMérieux), Architect (Abbott), and Liaison (DiaSorin) systems in 631 sera from pregnant women and newborns as well as anti-Toxoplasma IgG avidity with these three systems on 54 sera from pregnant women with positive IgG and IgM. The IgG and IgM results were in agreement in, respectively, 95.2% and 98.3% (Vidas versus Architect) and 96.9% and 95.3% (Vidas versus Liaison) of the samples. Specificities were excellent for all the assays, while Vidas sensitivities ranged (depending on the classification of gray zone results) from 93.8 to 98.4% for IgG (Architect, 84.4 to 93.8%; Liaison, 93.8%) and from 81.8 to 90.9% for IgM (Architect, 63.6%; Liaison, 81.8 to 90.9%). In seroconversion sequences, IgMs were generally detected simultaneously by the three assays, while Architect was the earliest assay to detect IgG. In noninfected children, maternally transmitted IgGs were detected for a longer time with Architect than with the other systems. IgMs were positive in only one infected child with the Vidas and Liaison systems. Significantly more sera were classified in the high-avidity category with Vidas than with Architect. This evaluation shows similar performances for Vidas and more recent systems. The Vidas system adequately detects toxoplasmosis in pregnant women and newborns. This system fits the needs of laboratories working on small routine series for first-line testing as well as expert laboratories, due to a high specificity and a powerful avidity test.

摘要

检测弓形虫病的血清学检测对于避免该疾病对新生儿可能产生的影响至关重要。本研究使用Vidas(生物梅里埃公司)、Architect(雅培公司)和Liaison(索灵公司)检测系统,对631份孕妇和新生儿血清进行了抗弓形虫IgG和IgM检测,并对54份IgG和IgM呈阳性的孕妇血清使用这三种检测系统进行了抗弓形虫IgG亲和力检测。IgG和IgM检测结果在样本中的一致性分别为:Vidas与Architect相比,IgG为95.2%,IgM为98.3%;Vidas与Liaison相比,IgG为96.9%,IgM为95.3%。所有检测方法的特异性均极佳,Vidas检测IgG的灵敏度范围(取决于灰区结果的分类)为93.8%至98.4%(Architect为84.4%至93.8%;Liaison为93.8%),检测IgM的灵敏度范围为81.8%至90.9%(Architect为63.6%;Liaison为81.8%至90.9%)。在血清转化序列中,三种检测方法通常能同时检测到IgM,而Architect是最早检测到IgG的方法。在未感染儿童中,与其他检测系统相比,Architect检测到母体传播IgG的时间更长。使用Vidas和Liaison检测系统时,只有一名感染儿童的IgM呈阳性。与Architect相比,Vidas将更多血清归类为高亲和力类别。该评估显示Vidas与更新的检测系统性能相似。Vidas检测系统能充分检测孕妇和新生儿中的弓形虫病。由于具有高特异性和强大的亲和力检测能力,该系统适合从事小规模常规检测的一线实验室以及专业实验室的需求。

相似文献

引用本文的文献

本文引用的文献

3
4
Toxoplasmosis: diagnosis, treatment, and prevention in congenitally exposed infants.弓形虫病:先天性感染婴儿的诊断、治疗和预防。
J Pediatr Health Care. 2011 Nov-Dec;25(6):355-64. doi: 10.1016/j.pedhc.2010.04.008. Epub 2010 Jun 12.
5
Clinical manifestations of ocular toxoplasmosis.眼弓形体病的临床表现。
Ocul Immunol Inflamm. 2011 Apr;19(2):91-102. doi: 10.3109/09273948.2011.564068.
10
Why prevent, diagnose and treat congenital toxoplasmosis?为什么要预防、诊断和治疗先天性弓形虫病?
Mem Inst Oswaldo Cruz. 2009 Mar;104(2):320-44. doi: 10.1590/s0074-02762009000200029.