Gilvin P J, Baker S T, Gibbens N J, Roberts G H, Tanner R J, Eakins J S, Hager L G, Daniels T J
Centre for Radiation, Chemical and Environmental Hazards, Public Health England, Chilton, Didcot OX11 0RQ, UK.
Radiat Prot Dosimetry. 2013 Dec;157(3):430-6. doi: 10.1093/rpd/nct144. Epub 2013 Jun 5.
A new head band dosemeter, for the measurement of eye lens dose in terms of Hp(3), has been type tested by Public Health England's Centre for Radiation, Environmental and Chemical Hazards [formerly part of the UK Health Protection Agency (HPA)]. The type tests were based on the International Standard ISO 12794, drawing also upon earlier work at HPA. The results show that, unlike many existing dosemeters, the new head band dosemeter correctly measures Hp(3) for beta radiations as well as photons.