This article describes the long-term evaluation of a nanoceria-polyacrylonitrile (CeO2-PAN) composite sorbent-based (68)Ge/(68)Ga generator reported. This generator used the new CeO2-PAN composite sorbent for preparation of the (68)Ge/(68)Ga generator. Since this sorbent has not been previously evaluated, a thorough long-term evaluation of the performance of the generator is necessary to ensure its applicability for clinical practice. The performance of the generator was evaluated in terms of (68)Ga yield, (68)Ge breakthrough, radioactive concentration of the (68)Ga solution, and suitability of the (68)Ga for the preparation of (68)Ga-labeled tracers. The (68)Ge/(68)Ga generator was able to provide a (68)Ga activity with consistent yields (>70%) and having acceptable radionuclidic (<10(-4)% of (68)Ge breakthrough), radiochemical, and chemical purities for an extended period of time. The eluted (68)GaCl3 is useful for the majority of the (68)Ga complexation chemistry.

摘要

本文描述了一种基于纳米氧化铈-聚丙烯腈(CeO2-PAN)复合吸附剂的(68)Ge/(68)Ga 发生器的长期评价。该发生器使用新型 CeO2-PAN 复合吸附剂制备(68)Ge/(68)Ga 发生器。由于该吸附剂尚未经过先前的评估,因此需要对发生器的性能进行彻底的长期评估,以确保其在临床实践中的适用性。从(68)Ga 产率、(68)Ge 穿透率、(68)Ga 溶液放射性浓度以及(68)Ga 用于制备(68)Ga 标记示踪剂的适用性等方面评价了发生器的性能。(68)Ge/(68)Ga 发生器能够在较长时间内提供具有一致产率(>70%)且放射性核素纯度可接受(<10-4%的(68)Ge 穿透率)的(68)Ga 活性,具有可接受的放射化学和化学纯度。洗脱的(68)GaCl3 适用于大多数(68)Ga 络合化学。

文献检索

告别复杂PubMed语法,用中文像聊天一样搜索,搜遍4000万医学文献。AI智能推荐,让科研检索更轻松。

立即免费搜索

文件翻译

保留排版,准确专业,支持PDF/Word/PPT等文件格式,支持 12+语言互译。

免费翻译文档

深度研究

AI帮你快速写综述,25分钟生成高质量综述,智能提取关键信息,辅助科研写作。

立即免费体验