Li Jun Yuan, Pow Edmond Ho Nang, Zheng Li Wu, Ma Li, Kwong Dora Lai Wan, Cheung Lim Kwong
Oral Rehabilitation, Prince Philip Dental Hospital, The University of Hong Kong, Hong Kong SAR, China.
Clin Oral Implants Res. 2014 Apr;25(4):506-10. doi: 10.1111/clr.12200. Epub 2013 Jun 9.
To investigate the impact of cover screw, resin embedment, and implant angulation on artifact of microcomputed tomography (micro-CT) scanning for implant.
A total of twelve implants were randomly divided into 4 groups: (i) implant only; (ii) implant with cover screw; (iii) implant with resin embedment; and (iv) implants with cover screw and resin embedment. Implants angulation at 0°, 45°, and 90° were scanned by micro-CT. Images were assessed, and the ratio of artifact volume to total volume (AV/TV) was calculated. A multiple regression analysis in stepwise model was used to determine the significance of different factors. One-way ANOVA was performed to identify which combination of factors could minimize the artifact.
In the regression analysis, implant angulation was identified as the best predictor for artifact among the factors (P < 0.001). Resin embedment also had significant effect on artifact volume (P = 0.028), while cover screw had not (P > 0.05). Non-embedded implants with the axis parallel to X-ray source of micro-CT produced minimal artifact.
Implant angulation and resin embedment affected the artifact volume of micro-CT scanning for implant, while cover screw did not.
研究覆盖螺丝、树脂包埋和种植体角度对种植体微计算机断层扫描(micro-CT)伪影的影响。
将12枚种植体随机分为4组:(i)仅种植体组;(ii)带覆盖螺丝的种植体组;(iii)树脂包埋的种植体组;(iv)带覆盖螺丝和树脂包埋的种植体组。通过micro-CT对角度为0°、45°和90°的种植体进行扫描。对图像进行评估,并计算伪影体积与总体积之比(AV/TV)。采用逐步模型的多元回归分析来确定不同因素的显著性。进行单因素方差分析以确定哪种因素组合可使伪影最小化。
在回归分析中,种植体角度被确定为各因素中伪影的最佳预测指标(P < 0.001)。树脂包埋对伪影体积也有显著影响(P = 0.028),而覆盖螺丝则无影响(P > 0.05)。轴与micro-CT X射线源平行的未包埋种植体产生的伪影最小。
种植体角度和树脂包埋会影响种植体micro-CT扫描的伪影体积,而覆盖螺丝则不会。