Türken Mehmet Aykut, Akdemir Mehmet, Uzun Bora, Ozkan Mustafa
Department of Orthopedics and Traumatology, Bolu İzzet Baysal State Hospital, Bolu, Turkey.
Acta Orthop Traumatol Turc. 2013;47(3):173-8. doi: 10.3944/aott.2013.2701.
The aim of this study was to compare the biomechanical strength of three fixation methods in humerus shaft fractures with medial butterfly fragment.
An AO 12B2 fracture with butterfly fragment was created in 21 synthetic cortical shell humeri. Humeri were divided into 3 groups which underwent bridge plating, antegrade intramedullary nailing or retrograde intramedullary nailing. Rotational and four-point bending test displacement curves were obtained.
Mean displacement values in the torsion tests were 37.38 (range: 25.75 to 55.69) mm in the bridge plating group, 26.55 (range: 21.25 to 41.81) mm in the antegrade nailing group, and 33.23 (range: 27.50 to 46.45) mm in the retrograde nailing group. Mean displacement values in the four-point bending test were 3.27 (range: 2.54 to 3.73) mm in the bridge plating group, 3.17 (range: 2.69 to 3.55) mm in the antegrade nailing group, and 3.15 (range: 2.10 to 4.03) mm in the retrograde nailing group. No significant difference was found among the groups (p>0.05).
The biomechanical stability of bridge plating, antegrade intramedullary nailing and retrograde intramedullary nailing appears to be similar in the fixation of humerus shaft fractures with medial butterfly fragment.
本研究旨在比较三种固定方法治疗伴有内侧蝶形骨块的肱骨干骨折的生物力学强度。
在21根合成皮质骨壳肱骨上制造带有蝶形骨块的AO 12B2骨折。将肱骨分为3组,分别接受桥接钢板固定、顺行髓内钉固定或逆行髓内钉固定。获得旋转和四点弯曲试验的位移曲线。
桥接钢板固定组扭转试验的平均位移值为37.38(范围:25.75至55.69)mm,顺行髓内钉固定组为26.55(范围:21.25至41.81)mm,逆行髓内钉固定组为33.23(范围:27.50至46.45)mm。桥接钢板固定组四点弯曲试验的平均位移值为3.27(范围:2.54至3.73)mm,顺行髓内钉固定组为3.17(范围:2.69至3.55)mm,逆行髓内钉固定组为3.15(范围:2.10至4.03)mm。各组之间未发现显著差异(p>0.05)。
在固定伴有内侧蝶形骨块的肱骨干骨折时,桥接钢板固定、顺行髓内钉固定和逆行髓内钉固定的生物力学稳定性似乎相似。