Kirk B W, Raber M B, Duke K R
J Appl Physiol. 1975 Jun;38(6):1140-2. doi: 10.1152/jappl.1975.38.6.1140.
A method to determine the P50 of whole blood is described using a modified American Optical reflectance oximeter, pump, and membrane tonometer, together with PO2, PCO2, and pH measurements in a standard blood gas machine. Determinations of P50 were made in 66 patients and normal subjects and in two situations where P50 was very low and very high. The results were compared to oxygen saturations calculated from measured oxygen content. The directly determined oxygen saturation agreed with the assumed saturation of 50 per cent in the oximeter within 0.5 per cent. The apparatus appears to be a simple and relatively inexpensive method to obtain P50 as long as blood carboxyhemoglobin or methemoglobin contents are not elevated.
描述了一种使用改良的美国光学反射式血氧计、泵和膜式眼压计,结合标准血气分析仪进行的PO2、PCO2和pH测量来测定全血P50的方法。对66例患者和正常受试者以及P50非常低和非常高的两种情况进行了P50测定。将结果与根据测得的氧含量计算出的氧饱和度进行比较。直接测定的氧饱和度与血氧计中假设的50%饱和度的偏差在0.5%以内。只要血液中碳氧血红蛋白或高铁血红蛋白含量不升高,该仪器似乎是一种获取P50的简单且相对便宜的方法。