Krishna A Chaitanya, Sathiyaraj M, Saravanan R S, Chelladurai R, Vignesh R
Department of Bioanalaysis, Bombay Bioresearch Center, Govandi, Mumbai-400 043, India.
Indian J Pharm Sci. 2012 Nov;74(6):541-8. doi: 10.4103/0250-474X.110599.
A simple, rapid, specific and sensitive liquid chromatography tandem mass spectrometric method has been developed and validated for the determination of doxycycline from the human plasma. Doxycycline is extracted from human plasma by solid phase extraction. Demeclocycline was used as an internal standard. Detection was performed at transitions of 444.800→428.200 for doxycycline and 464.700→448.100 for demeclocycline using mass spectrometry. Chromatographic separation of analyte and internal standard were carried out using a reverse phase C18, column at 0.500 ml/min flow. The assay of doxycycline is linear over the range of 0.055-7.612 μg/ml, with a precision <14.83%, regression coefficient (r(2))=0.9961 and the limit of quantification in plasma for doxycycline was 0.055 μg/ml. Mean extraction recovery obtained was 95.55%. Samples are stable at room temperature for 6 h, processed samples were stable at least for 30.20 h and also stable at three freeze-thaw cycles. The method has been used to perform pharmacokinetic and bioequivalence studies in human plasma.
已开发并验证了一种简单、快速、特异且灵敏的液相色谱串联质谱法,用于测定人血浆中的多西环素。多西环素通过固相萃取从人血浆中提取。地美环素用作内标。使用质谱法在多西环素的444.800→428.200和地美环素的464.700→448.100跃迁处进行检测。使用反相C18柱以0.500 ml/min的流速对分析物和内标进行色谱分离。多西环素的测定在0.055 - 7.612 μg/ml范围内呈线性,精密度<14.83%,回归系数(r(2))=0.9961,多西环素在血浆中的定量限为0.055 μg/ml。获得的平均提取回收率为95.55%。样品在室温下稳定6小时,处理后的样品至少稳定30.20小时,并且在三个冻融循环中也稳定。该方法已用于进行人血浆中的药代动力学和生物等效性研究。