Guy J M, Legg E F
Department of Clinical Chemistry, East Birmingham Hospital, Bordesley Green East, UK.
Ann Clin Biochem. 1990 May;27 ( Pt 3):223-6. doi: 10.1177/000456329002700307.
An improved HPLC method for the measurement of serum creatinine is described. Separation is effected using a strong cation exchange column at pH 4.7. Analytical recovery, precision and linearity are satisfactory, and the method correlates well with a kinetic Jaffé procedure. The method provides a means of accurately measuring creatinine and may be of use in the investigation of interference in creatinine assays.
描述了一种改进的用于测定血清肌酐的高效液相色谱法。使用强阳离子交换柱在pH 4.7条件下进行分离。分析回收率、精密度和线性均令人满意,且该方法与动力学Jaffé法相关性良好。该方法提供了一种准确测量肌酐的手段,可用于研究肌酐测定中的干扰情况。