Goresky C A, Gordon E R
McGill University Medical Clinic, Montreal, Quebec, Canada.
J Chromatogr. 1990 Jun 8;528(1):123-41. doi: 10.1016/s0378-4347(00)82368-3.
A fast, sensitive high-performance liquid chromatographic method has been developed for the separation and quantitation of biliary bile pigments; this utilizes a C18 reversed-phase column with two solvents, a buffer and an organic solvent, which were changed in a linear gradient from a polar to a less polar combination. Nine glycosidic conjugates of bilirubin as well as unconjugated bilirubin and a suitable internal standard, unconjugated mesobilirubin IX alpha, were all separated to baseline by gradient elution; the species eluted in a polar to less polar fashion. Increasing the molarity of the solvent decreased the binding of non-glucuronide pigments to the column, with a decrease in their retention times, whereas for bilirubin monoglucuronide they increased. Decrease in pH, similarly, preferentially increased bilirubin monoglucuronide retention times.